KR20170094758A - Method of Manufacturing Composition for Catheter - Google Patents
Method of Manufacturing Composition for Catheter Download PDFInfo
- Publication number
- KR20170094758A KR20170094758A KR1020170013679A KR20170013679A KR20170094758A KR 20170094758 A KR20170094758 A KR 20170094758A KR 1020170013679 A KR1020170013679 A KR 1020170013679A KR 20170013679 A KR20170013679 A KR 20170013679A KR 20170094758 A KR20170094758 A KR 20170094758A
- Authority
- KR
- South Korea
- Prior art keywords
- catheter
- poly
- carbon nanotube
- zinc oxide
- zno
- Prior art date
Links
Images
Classifications
-
- C—CHEMISTRY; METALLURGY
- C08—ORGANIC MACROMOLECULAR COMPOUNDS; THEIR PREPARATION OR CHEMICAL WORKING-UP; COMPOSITIONS BASED THEREON
- C08K—Use of inorganic or non-macromolecular organic substances as compounding ingredients
- C08K3/00—Use of inorganic substances as compounding ingredients
- C08K3/02—Elements
- C08K3/04—Carbon
- C08K3/041—Carbon nanotubes
-
- C—CHEMISTRY; METALLURGY
- C08—ORGANIC MACROMOLECULAR COMPOUNDS; THEIR PREPARATION OR CHEMICAL WORKING-UP; COMPOSITIONS BASED THEREON
- C08J—WORKING-UP; GENERAL PROCESSES OF COMPOUNDING; AFTER-TREATMENT NOT COVERED BY SUBCLASSES C08B, C08C, C08F, C08G or C08H
- C08J3/00—Processes of treating or compounding macromolecular substances
- C08J3/20—Compounding polymers with additives, e.g. colouring
-
- A—HUMAN NECESSITIES
- A61—MEDICAL OR VETERINARY SCIENCE; HYGIENE
- A61L—METHODS OR APPARATUS FOR STERILISING MATERIALS OR OBJECTS IN GENERAL; DISINFECTION, STERILISATION OR DEODORISATION OF AIR; CHEMICAL ASPECTS OF BANDAGES, DRESSINGS, ABSORBENT PADS OR SURGICAL ARTICLES; MATERIALS FOR BANDAGES, DRESSINGS, ABSORBENT PADS OR SURGICAL ARTICLES
- A61L29/00—Materials for catheters, medical tubing, cannulae, or endoscopes or for coating catheters
- A61L29/04—Macromolecular materials
- A61L29/06—Macromolecular materials obtained otherwise than by reactions only involving carbon-to-carbon unsaturated bonds
-
- C—CHEMISTRY; METALLURGY
- C07—ORGANIC CHEMISTRY
- C07F—ACYCLIC, CARBOCYCLIC OR HETEROCYCLIC COMPOUNDS CONTAINING ELEMENTS OTHER THAN CARBON, HYDROGEN, HALOGEN, OXYGEN, NITROGEN, SULFUR, SELENIUM OR TELLURIUM
- C07F7/00—Compounds containing elements of Groups 4 or 14 of the Periodic System
- C07F7/02—Silicon compounds
- C07F7/08—Compounds having one or more C—Si linkages
- C07F7/0834—Compounds having one or more O-Si linkage
-
- C—CHEMISTRY; METALLURGY
- C08—ORGANIC MACROMOLECULAR COMPOUNDS; THEIR PREPARATION OR CHEMICAL WORKING-UP; COMPOSITIONS BASED THEREON
- C08G—MACROMOLECULAR COMPOUNDS OBTAINED OTHERWISE THAN BY REACTIONS ONLY INVOLVING UNSATURATED CARBON-TO-CARBON BONDS
- C08G77/00—Macromolecular compounds obtained by reactions forming a linkage containing silicon with or without sulfur, nitrogen, oxygen or carbon in the main chain of the macromolecule
- C08G77/04—Polysiloxanes
-
- C08K3/005—
-
- C—CHEMISTRY; METALLURGY
- C08—ORGANIC MACROMOLECULAR COMPOUNDS; THEIR PREPARATION OR CHEMICAL WORKING-UP; COMPOSITIONS BASED THEREON
- C08K—Use of inorganic or non-macromolecular organic substances as compounding ingredients
- C08K3/00—Use of inorganic substances as compounding ingredients
- C08K3/02—Elements
- C08K3/04—Carbon
-
- C—CHEMISTRY; METALLURGY
- C08—ORGANIC MACROMOLECULAR COMPOUNDS; THEIR PREPARATION OR CHEMICAL WORKING-UP; COMPOSITIONS BASED THEREON
- C08K—Use of inorganic or non-macromolecular organic substances as compounding ingredients
- C08K3/00—Use of inorganic substances as compounding ingredients
- C08K3/18—Oxygen-containing compounds, e.g. metal carbonyls
- C08K3/20—Oxides; Hydroxides
- C08K3/22—Oxides; Hydroxides of metals
-
- C—CHEMISTRY; METALLURGY
- C08—ORGANIC MACROMOLECULAR COMPOUNDS; THEIR PREPARATION OR CHEMICAL WORKING-UP; COMPOSITIONS BASED THEREON
- C08L—COMPOSITIONS OF MACROMOLECULAR COMPOUNDS
- C08L83/00—Compositions of macromolecular compounds obtained by reactions forming in the main chain of the macromolecule a linkage containing silicon with or without sulfur, nitrogen, oxygen or carbon only; Compositions of derivatives of such polymers
- C08L83/04—Polysiloxanes
-
- C—CHEMISTRY; METALLURGY
- C08—ORGANIC MACROMOLECULAR COMPOUNDS; THEIR PREPARATION OR CHEMICAL WORKING-UP; COMPOSITIONS BASED THEREON
- C08K—Use of inorganic or non-macromolecular organic substances as compounding ingredients
- C08K3/00—Use of inorganic substances as compounding ingredients
- C08K3/18—Oxygen-containing compounds, e.g. metal carbonyls
- C08K3/20—Oxides; Hydroxides
- C08K3/22—Oxides; Hydroxides of metals
- C08K2003/2296—Oxides; Hydroxides of metals of zinc
-
- C—CHEMISTRY; METALLURGY
- C08—ORGANIC MACROMOLECULAR COMPOUNDS; THEIR PREPARATION OR CHEMICAL WORKING-UP; COMPOSITIONS BASED THEREON
- C08K—Use of inorganic or non-macromolecular organic substances as compounding ingredients
- C08K2201/00—Specific properties of additives
- C08K2201/011—Nanostructured additives
Abstract
Description
본 발명은 폴리 카테터에 사용되는 조성물을 제조하는 방법에 관한 것이다.The present invention relates to a method of making a composition for use in a poly catheter.
일반적으로 뇌졸증과 같은 대뇌질환이나 척추손상 등에 의해 일어나는 전신 또는 하반신 마비환자는 고령인구의 증가와 교통사고 또는 산업재해의 급증 등의 원인에 의해 해마다 증가되고 있는 실정이다.In general, patients with systemic or paraplegic paralysis caused by cerebral diseases such as stroke or spinal injuries are increasing year by year due to an increase in the elderly population and a rapid increase in traffic accidents or industrial accidents.
위와 같은 환자에서 필연적으로 방광 마비가 동반되는데, 방광마비에 대한 치료는 환자의 예후에 전적으로 좌우하게 되며, 이러한 환자에 대한 치료법으로 폴리 카테터(Foley Catheter)를 방광 내에 유지시켜 소변을 외부로 배출하게 된다.In this patient, bladder paralysis is inevitably accompanied by treatment of the bladder paralysis, which is entirely dependent on the prognosis of the patient, and a polycatheter (Foley Catheter) is maintained in the bladder as a treatment for these patients, do.
폴리 카테터는 튜브형 카테터 본체의 말단 부분에 폴리(Foley)를 접합시켜, 외부로부터 유입되는 유체에 의해 폴리가 팽창되어 풍선(Ballon) 형상을 가지도록 함으로써, 카테터가 방광내에 유지되도록 한다.The poly-catheter is configured such that poly (Foley) is attached to the distal end portion of the tubular catheter body so that the poly is inflated by the fluid introduced from the outside to have a balloon shape so that the catheter is held in the bladder.
종래의 항생 카테터의 경우, 실리콘 재질의 폴리 카테터에 항생 약품 또는 물질을 도포하여 세균의 침입을 억제하여 왔다. 이는 초기에 항생제로 인해 항생효과가 발생하나, 도뇨관의 특성상 요로에 2~3일 이상을 삽관하고 있어 생물막 생성(biofilm forming)이 필연적으로 발생하게 된다. In the case of a conventional antibiotic catheter, an antibiotic drug or a substance is applied to a poly catheter made of a silicone to suppress invasion of bacteria. Although antibiotics are initially effective in antibiotics, biofilm formation is inevitable due to intubation of the urinary tract for 2 to 3 days due to the nature of the catheter.
이와 같은 생물막의 생성으로 인해 폴리 카테터의 항생 효과가 떨어지거나 사라져, 요로감염과 같은 합병증이 발생하게 되며, 이를 위한 처치가 수반되어 입원기간이 길어지는 문제점이 있다.Such biofilm formation causes the antibiotic effect of the poly catheter to drop or disappear, resulting in complications such as urinary tract infection, and the procedure is accompanied with a problem that the length of hospitalization is prolonged.
또한, 종래의 폴리 카테터는 표면에 도포한 항생 약품 또는 물질이 항시 체결 내에 유치되기 때문에 이로 인한 요로감염, 결석형성 등으로 전체환자의 약40%에서 신부전이 초래되어 이러한 환자의 가장 큰 사망원인이 되었다. In addition, since conventional anticancer drugs or substances applied to the surface of the conventional catheter catheter are always kept in the joint, urinary tract infections and stones are formed, resulting in renal failure in about 40% of all patients. .
한편, 위와 같은 항생 카테터의 문제점을 해결하고자 카테터에 금, 은 또는 은나노 등과 같은 항균 물질을 코팅한 제품들이 있으나, 일정 기간이상 사용하는 경우 코팅된 항균 물질이 박리되어 항균성이 감소되는 문제가 있었다.On the other hand, in order to solve the problem of the above-mentioned antibiotic catheter, there is a product in which a catheter is coated with an antimicrobial substance such as gold, silver or silver nano. However, when the antimicrobial substance is used over a certain period of time,
상기한 종래 기술의 문제점 및 과제에 대한 인식은 본 발명의 기술 분야에서 통상의 지식을 가진 자에게 자명한 것이 아니므로 이러한 인식을 기반으로 선행기술들과 대비한 본 발명의 진보성을 판단하여서는 아니 됨을 밝혀둔다.The recognition of the problems and problems of the prior art is not obvious to a person having ordinary skill in the art, so that the inventive step of the present invention should not be judged based on the recognition based on such recognition I will reveal.
본 발명은 향상된 향균성을 가지며 장기간의 사용에도 향균성이 유지될 수 있도록 하는 폴리 카테터에 사용되는 조성물의 제조 방법을 제공하는 것을 목적으로 한다.It is an object of the present invention to provide a method for producing a composition for use in a poly catheter, which has improved antibacterial properties and is capable of maintaining antibacterial property even after prolonged use.
본 발명의 일실시예에 따른 폴리 카테터용 조성물 제조 방법은, 분산된 탄소나노튜브와 징크옥사이드(ZnO)를 실리콘에 복합 처방하여 탄소나노튜브 중합체(CNT Polymer)가 실리콘에 배합된 소재를 구성한다.The method for preparing a composition for a poly catheter according to an embodiment of the present invention comprises dispersing carbon nanotubes and zinc oxide (ZnO) in combination with silicon to form a material in which a carbon nanotube polymer (CNT polymer) is mixed with silicon .
본 발명의 일실시예에 따르면, 탄소나노튜브와 징크옥사이드(ZnO)가 결합된 탄소나노튜브 중합체(CNT Polymer)가 실리콘에 배합된 소재로 카테터 본체와 카테터용 폴리를 구성함으로써, 별도의 항생 물질의 도포 또는 코팅 없이 세균 감염의 근윈인 생물막의 형성을 억제할 수 있다.According to an embodiment of the present invention, a catheter body and a catheter poly are made of a material in which carbon nanotube (CNT) polymer combined with zinc oxide (CNT) is mixed with silicon, It is possible to inhibit the formation of a biofilm, which is the root of bacterial infection, without application or coating.
또한, 탄소나노튜브 중합체가 가진 생체전위효과 유도를 폴리에 균일하게 유지하여 폴리에 협착하는 세균을 무활동성으로 변이시켜 항세균성을 유지할 수 있는 효과를 지니고 있다.In addition, it has an effect of maintaining antibacterial property by mutating inactivating bacteria of poly-stenosis by maintaining induction effect of carbon nanotube polymer with bio-potential uniformly in poly.
또한, 본 발명에 따른 카테터용 폴리는 항생물질의 내성과 같은 부작용이 없고, 탄소나노튜브 중합체가 가지고 있는 정전량에 따라 그 수명이 정해지는 특성과 탄소나노튜브의 특성인 높은 열전도로 인하여, 인체 삽입 과정에서 환자의 거부감을 최소화 하고 추가적인 교체 비용의 절감과 감염에 따른 의료비용증가를 막아주는 효과를 가지고 있다.The catheter poly according to the present invention has no side effects such as antibiotic resistance and is characterized in that the lifetime of the carbon nanotube polymer is determined by the amount of electrostatic capacitance possessed by the carbon nanotube polymer, It has the effect of minimizing the rejection of the patient during the insertion process, reducing the additional replacement cost and preventing the increase of medical expenses due to the infection.
상기한 본 발명의 효과는 본 발명에 따른 다양한 효과 중 하나일 뿐이며, 본 발명은 실시예 적용방식에 따라 다양한 형태로 실현될 수 있다.The effects of the present invention described above are only one of various effects according to the present invention, and the present invention can be realized in various forms according to the application method of the embodiments.
도 1은 본 발명의 일실시예에 따른 폴리 카테터의 구성을 나타내는 사시도이다.
도 2는 도 1에 도시된 폴리 카테터의 단면 구성에 대한 일예를 나타내는 단면도이다.
도 3 내지 도 8은 폴리 카테터와 카테터용 폴리를 구성하는 소재의 생물막 형성 억제 효과에 대한 실험 결과들을 나타내는 도면들이다.
도 9는 폴리 카테터와 카테터용 폴리를 구성하는 소재의 인체 삽입시 이물감 감소 효과를 설명하기 위한 실험 결과를 나타내는 도면이다.
도 10은 본 발명의 일실시예에 따른 카테터용 폴리의 구성 및 동작을 설명하기 위한 도면이다.
도 11은 본 발명에 따른 카테터용 폴리의 구성에 대한 일실시예를 보다 상세히 나타내는 도면이다.
도 12 내지 도 16은 도 11에 도시된 카테터용 폴리의 단면 구성에 대한 예들을 나타내는 단면도들이다.1 is a perspective view showing a configuration of a poly catheter according to an embodiment of the present invention.
2 is a cross-sectional view showing an example of a sectional configuration of the polycatheter shown in Fig.
FIGS. 3 to 8 are diagrams showing experimental results on biofilm formation inhibitory effect of the material constituting the poly catheter and the catheter poly.
Fig. 9 is a view showing experimental results for explaining the effect of reducing the foreign body sensation when a human body is inserted into a poly catheter and a catheter poly.
10 is a view for explaining the configuration and operation of a catheter poly according to an embodiment of the present invention.
Figure 11 is a more detailed view of one embodiment of the configuration of a catheter poly according to the present invention.
Figs. 12 to 16 are cross-sectional views showing examples of the cross-sectional configuration of the catheter poly shown in Fig.
이하, 첨부된 도면들을 참조하여 본 발명의 실시예에 따른 폴리 카테터, 카테터용 폴리 및 그들의 제조 또는 형성 방법들에 관하여 상세히 설명한다.DETAILED DESCRIPTION OF THE PREFERRED EMBODIMENTS Reference will now be made in detail to the preferred embodiments of the present invention, examples of which are illustrated in the accompanying drawings.
본 발명의 상술한 목적, 특징들 및 장점은 첨부된 도면과 관련된 다음의 상세한 설명을 통하여 보다 분명해질 것이다. 이하 첨부된 도면을 참조하여 본 발명에 따른 바람직한 실시예들을 상세히 설명한다. 명세서 전체에 걸쳐서 동일한 참조번호들은 동일한 구성요소들을 나타낸다. 또한, 본 발명과 관련된 공지 기능 혹은 구성에 대한 구체적인 설명이 본 발명의 요지를 불필요하게 흐릴 수 있다고 판단되는 경우, 그 상세한 설명을 생략한다.The foregoing objects, features and advantages of the present invention will become more apparent from the following detailed description taken in conjunction with the accompanying drawings. DETAILED DESCRIPTION OF THE PREFERRED EMBODIMENTS Reference will now be made in detail to the preferred embodiments of the present invention, examples of which are illustrated in the accompanying drawings. Like reference numerals designate like elements throughout the specification. In the following description, well-known functions or constructions are not described in detail since they would obscure the invention in unnecessary detail.
도 1은 본 발명의 일실시예에 따른 폴리 카테터의 구성을 사시도로 도시한 것이며, 도 2는 폴리 카테터의 단면 구성에 대한 일예를 도시한 것으로 도 1에 도시된 B-B선에 따른 횡단면도를 나타낸 것이다.FIG. 1 is a perspective view of a configuration of a poly catheter according to an embodiment of the present invention, and FIG. 2 is a cross-sectional view of the poly catheter taken along line BB of FIG. 1 .
본 발명의 일실시예에 따른 카테터는 환자의 방광 내의 소변을 배출하기 위한 요도 카테터(Urine catheter)일 수 있으며, 카테터 본체의 일측에 소변유입구와 풍선 형태의 팽창 가능한 폴리가 설치되고, 카테터 본체의 중심에 위치한 주관이 본체의 타측에 위치한 소변배출부와 연통되어 환자의 방광 내의 소변을 배출하는 구조를 갖는 폴리(Foley) 카테터일 수 있다.A catheter according to an embodiment of the present invention may be a urine catheter for discharging urine in a bladder of a patient and includes a urine inlet and an inflatable poly in the form of a balloon at one side of the catheter body, And a main pipe positioned at the center of the main body communicates with a urine discharging part located on the other side of the main body to discharge urine in the bladder of the patient.
다만, 본 발명의 일실시예에 따른 카테터 및 카테터용 폴리는 상기한 바와 같은 요도 카테터 이외에, 심혈관 카테터 등 다양한 카테터에 적용될 수도 있다.However, the catheter and catheter poly according to an embodiment of the present invention may be applied to various catheters such as a cardiovascular catheter, in addition to the urethral catheter as described above.
도 1 및 도 2를 참조하면, 폴리 카테터는 튜브형 관체인 카테터 본체(100)와, 외부로부터 유입되는 유체에 의해 팽창 가능하도록 카테터 본체(100)에 접합되는 카테터용 폴리(200)를 포함하여 구성될 수 있다.Referring to Figures 1 and 2, a polycatheter comprises a
보다 구체적으로, 카테터 본체(100)는 일단(50)이 막혀 있고, 소변이 이동하는 소변통로(110)와 유체가 이동하는 유체통로(120, 121)가 내부에 각각 형성되어 있을 수 있다.More specifically, the catheter
또한, 소변유입구(11)는 카테터 본체(100)의 소변통로(110)와 연결되어 형성되며, 유체유출구(21)는 카테터 본체(100)의 유체통로(120, 121)와 연결될 수 있다.The
도 2에서는, 카테터 본체(100)에 하나의 소변통로(110)와 두 개의 유체통로들(120, 121)이 형성된 것을 예로 들어 본 발명의 일실시예에 따른 카테터 및 카테터용 폴리를 설명하였으나, 본 발명은 이에 한정되지 아니한다.2, a catheter and a catheter for a catheter according to an embodiment of the present invention have been described by taking as an example a
예를 들어, 카테터 본체(100)에 하나의 유체통로가 형성되어 있을 수도 있으며, 도 1에 도시된 바와 같은 투 웨이(Two-Way) 방식의 카테터가 아닌 쓰리 웨이(Three-Way)의 방식의 카테터의 경우 소변통로(110)와는 별도로 약품이 투입되기 위한 약품통로(미도시)가 추가로 구비될 수 있다.For example, one fluid passageway may be formed in the
상기한 바와 같은 유체통로(120, 121)는 각각 유체유출구(21)와 연결되어, 유체유입부(23)를 통해 외부로부터 유입되는 유체가 유체통로(120, 121) 및 유체유출구(21)를 통해 카테터용 폴리(200)가 접합된 부분까지 전달되도록 할 수 있다.The
여기서, 상기 유체는 공기 등과 같은 기체이거나 또는 식염수 등과 같은 액체일 수 있다.Here, the fluid may be a gas such as air or a liquid such as saline.
카테터용 폴리(200)는 유체가 유입됨에 따라 신축 또는 팽창이 가능한 소재로 구성되며, 카테터 본체(100)에 형성된 유체유출구(21)를 감싸면서 카테터 본체(100)에 접합될 수 있다.The poly (200) for catheter is composed of a material which can expand or contract as the fluid flows, and can be bonded to the catheter body (100) while surrounding the fluid outlet (21) formed in the catheter body (100).
도 1 및 도 2를 참조하여 설명한 바와 같은 구성을 가지는 본 발명의 일실시예에 따른 폴리 카테터 및 카테터용 폴리는, 실리콘에 탄소나노튜브 중합체(CNT Polymer)가 배합된 소재로 구성되어, 별도의 항생 물질의 도포 또는 코팅 없이 세균 감염의 근윈인 생물막의 형성을 억제할 수 있다.The poly catheter and the catheter poly according to an embodiment of the present invention having the configurations described with reference to Figs. 1 and 2 are made of a material in which carbon nanotube polymer (CNT polymer) is mixed with silicone, It is possible to inhibit the formation of a biofilm, which is the root of bacterial infection, without application or coating of antibiotics.
탄소나노튜브(carbon nanotube, CNT)는 탄소원자가 만드는 원통형의 결정으로 직경 2~20nm(1nm는 1/1,000,000m), 길이는 수백~수천nm로 이루어진다. 탄소나노튜브에서 탄소원자 하나는 주위에 다른 탄소원자 3개와 sp2결합을 하여 육각형 벌집무늬를 형성하며, 이 튜브의 직경이 대략 수 나노미터(nanometer, nm) 정도로 극히 작기 때문에 나노튜브라고 부르게 되었다.Carbon nanotubes (CNTs) are cylindrical crystals made of a carbon source and have a diameter of 2 to 20 nm (1 nm to 1 / 1,000,000 m) and a length of several hundred to several thousand nm. One carbon atom in a carbon nanotube forms a hexagonal honeycomb pattern by sp2 bonding with three other carbon atoms around it, which is called a nanotube because the diameter of the tube is very small, about nanometers (nm).
또한, 탄소나노튜브 중합체(CNT Polymer)는 탄소나노튜브(CNT)와 징크옥사이드(ZnO)가 결합된 중합체이며, 탄소나노튜브(CNT)와 징크옥사이드(ZnO)가 서로 동일한 비율로 중합되거나, 또는 징크옥사이드(ZnO)가 탄소나노튜브(CNT) 보다 높은 비율로 중합될 수 있으며, 필요에 따라 그 반대의 경우도 가능할 수 있다.The carbon nanotube polymer (CNT polymer) is a polymer in which carbon nanotubes (CNT) and zinc oxide (ZnO) are bonded, and carbon nanotubes (CNT) and zinc oxide (ZnO) Zinc oxide (ZnO) can be polymerized at a higher ratio than carbon nanotubes (CNT), and vice versa, if necessary.
본 발명의 일실시예에 따른 카테터를 구성하는 카테터 본체(100)는 실리콘 100 중량부 대비 탄소나노튜브 중합체 1.0 내지 2.2 중량부로 배합된 소재로 구성될 수 있으나, 상기 배합비율은 카테터의 기능성에 따라 변동 가능할 수 있다.The
본 발명에 따르면, 카테터 본체(100)의 구성물인 탄소나노튜브 중합체는 삽관된 인체 내 전위에 반응하여 일정한 정전용량을 가지게 됨으로써, 이러한 정전용량은 인체에는 무해하나 세균과 생물막에는 치명적인 정전효과(Galvanic effect)를 가지게 되어 생물막 형성을 최소화 할 수 있도록 하며, 탄소나노튜브의 특성인 높은 Thermal Conductivity로 인하여 인체 삽입 과정에서 해당자의 거부감을 최소화할 수 있다.According to the present invention, the carbon nanotube polymer as a constituent of the
또한, 기존 항생 물질을 도포 또는 코팅한 카테터의 경우 생물막 형성과 세균 감염으로 인해 1주 이상 사용이 불가하나, 본 발명에 따른 폴리 카테터는 상기한 바와 같은 효과를 가지는 탄소나노튜브 중합체(CNT ploymer)가 실리콘에 내첨됨에 따라 최소 4~5주의 사용연한을 가지게 될 수 있다.In the case of a catheter coated or coated with a conventional antibiotic, it is impossible to use the catheter for more than one week due to biofilm formation and bacterial infection. However, the poly catheter according to the present invention is a CNT ploymer having the above- Can be used for at least 4 to 5 weeks due to the presence of silicon.
본 발명의 일실시예에 따르면, 상기 탄소나노튜브(CNT)는 다중벽 탄소나노튜브(Multi-Walled Carbon Nanotube, MWNT)일 수 있으며, 이는 다중벽 탄소나노튜브(MWNT)가 고체 상태에서의 사용에 장점이 있으며 가격 측면에서 상용화 가능성이 높기 때문이다.According to an embodiment of the present invention, the carbon nanotube (CNT) may be a multi-walled carbon nanotube (MWNT), which may be a multi-walled carbon nanotube (MWNT) And it is likely to be commercialized in terms of price.
상기한 바와 같이 본 발명의 일실시예에 따른 폴리 카테터와 카테터용 폴리는 각각 아래의 화학식 1과 같이 탄소나노튜브(CNT)와 징크옥사이드(ZnO)가 결합된 탄소나노튜브 중합체(CNT Polymer)가 실리콘에 배합된 소재로 구성될 수 있다.As described above, the poly catheter and the catheter poly according to an embodiment of the present invention include a carbon nanotube polymer (CNT polymer) having carbon nanotubes (CNT) and zinc oxide (ZnO) It can be composed of material blended with silicone.
상기 화학식 1에서 m, n, p는 실리콘, 징크옥사이드(ZnO), 탄소나노튜브(CNT)의 분자수를 각각 나타내는 것으로, m은 50 내지 300, n은 7 내지 30, p는 10 내지 50의 범위를 가질 수 있으나, 본 발명은 이에 한정하지 않는다.Wherein m, n and p represent the number of molecules of silicon, zinc oxide (ZnO) and carbon nanotubes (CNT), m is from 50 to 300, n is from 7 to 30, p is from 10 to 50 But the present invention is not limited thereto.
한편, 카테터와 카테터용 폴리에 있어서, 상기 m, n, p는 서로 상이하게 설정될 수 있다.On the other hand, in the catheter and the catheter poly, the above-mentioned m, n, and p may be set to be different from each other.
또한, 본 발명의 일실시예에 따른 카테터를 구성하기 위한 카테터 본체(100)와 카테터용 폴리(200)는 각각, 탄소나노튜브(CNT)와 징크옥사이드(ZnO)가 결합된 탄소나노튜브 중합체(CNT Polymer)와 실리콘이 일정 비율로 배합된 젤상의 소재를 압출하여 튜브형의 관체로 형성된 것일 수 있다.The
여기서, 상기 소재는 화학기상증착법(CVD composite)을 이용하여 분산된 탄소나노튜브(CNT)와 징크옥사이드(ZnO)를 실리콘에 복합 처방하여 구성될 수 있으며, 예를 들어, 1,000,000 Pa의 압력 및 50℃의 온도 조건에서 약 30분의 분산공정을 통해 제조될 수 있다.Here, the material may be formed by compounding carbon nanotubes (CNT) and zinc oxide (ZnO) dispersed using a CVD composite in silicon. For example, a pressure of 1,000,000 Pa and a pressure of 50 Lt; 0 > C under a temperature of about 30 minutes.
그에 따라, 탄소나노튜브(CNT)와 징크옥사이드(ZnO)로 이루어진 탄소나노튜브 중합체가 실리콘에 균일하에 내첨될 수 있으며, 그에 따라 해당 소재로 제조된 카테터 및 카테터용 폴리는 위치에 관계없이 균일한 항균력을 가질 수 있다.Accordingly, a carbon nanotube polymer composed of carbon nanotubes (CNT) and zinc oxide (ZnO) can be uniformly adhered to silicon, so that the catheter made of the material and the poly for the catheter can be uniformly It can have antibacterial activity.
이하, 도 3 내지 도 8을 참조하여 본 발명의 일실시예에 따른 폴리 카테터와 카테터용 폴리를 구성하는 소재의 생물막 형성 억제 효과에 대해 설명하기로 한다.Hereinafter, the biofilm formation inhibiting effect of the material constituting the poly catheter and the poly for the catheter according to one embodiment of the present invention will be described with reference to FIGS. 3 to 8. FIG.
도 3 내지 도 5는 상기한 바와 같은 소재로 구성된 카테터 절편에 요로감염 주요 균인 E. Coli.(Escherichia Coli.)을 각각 3일, 5일, 7일 동안 배양시킨 후, CV(Crystal Violet) 방법을 이용해 생물막의 형성 정도를 실험한 결과이다.3 to 5 show that Escherichia coli, a major strain of urinary tract infection, was cultured on a catheter slice composed of the above materials for 3 days, 5 days, and 7 days, respectively, and then CV (Crystal Violet) The biofilm formation rate was measured by using an optical microscope.
도 3을 참조하면, E. Coli.(Escherichia Coli.)을 3일 동안 배양시킨 후, 징크옥사이드(ZnO)의 배합비율이 0%인 경우에는 흡광도의 평균값이 0.303으로 측정되었고, 징크옥사이드(ZnO)의 배합비율이 1%인 경우에는 흡광도의 평균값이 0.326으로 측정되었으며, 징크옥사이드(ZnO)의 배합비율이 2%인 경우에는 흡광도의 평균값이 0.252로 측정되었고, 징크옥사이드(ZnO)의 배합비율이 3%인 경우에는 흡광도의 평균값이 0.299로 측정되었다.3, when E. coli. (Escherichia coli.) Was cultured for 3 days, the average value of the absorbance was 0.303 when the blending ratio of zinc oxide (ZnO) was 0%, and the average value of zinc oxide ) Was 1%, the average value of absorbance was 0.326. When the compounding ratio of zinc oxide (ZnO) was 2%, the average value of absorbance was 0.252, and the mixing ratio of zinc oxide (ZnO) In case of 3%, the average value of absorbance was measured as 0.299.
도 4를 참조하면, E. Coli.(Escherichia Coli.)을 5일 동안 배양시킨 후, 징크옥사이드(ZnO)의 배합비율이 0%인 경우에는 흡광도의 평균값이 0.362로 측정되었고, 징크옥사이드(ZnO)의 배합비율이 1%인 경우에는 흡광도의 평균값이 0.380으로 측정되었으며, 징크옥사이드(ZnO)의 배합비율이 2%인 경우에는 흡광도의 평균값이 0.356로 측정되었고, 징크옥사이드(ZnO)의 배합비율이 3%인 경우에는 흡광도의 평균값이 0.448로 측정되었다.4, when E. coli. (Escherichia coli.) Was cultured for 5 days, the average value of absorbance was 0.362 when the blending ratio of zinc oxide (ZnO) was 0%, and the average value of zinc oxide ) Was 1%, the average value of the absorbance was measured to be 0.380. When the compounding ratio of zinc oxide (ZnO) was 2%, the average value of the absorbance was 0.356 and the mixing ratio of zinc oxide (ZnO) In case of 3%, the average value of absorbance was measured as 0.448.
도 5를 참조하면, E. Coli.(Escherichia Coli.)을 7일 동안 배양시킨 후, 징크옥사이드(ZnO)의 배합비율이 0%인 경우에는 흡광도의 평균값이 0.486으로 측정되었고, 징크옥사이드(ZnO)의 배합비율이 1%인 경우에는 흡광도의 평균값이 0.425로 측정되었으며, 징크옥사이드(ZnO)의 배합비율이 2%인 경우에는 흡광도의 평균값이 0.407로 측정되었고, 징크옥사이드(ZnO)의 배합비율이 3%인 경우에는 흡광도의 평균값이 0.413으로 측정되었다.5, when E. coli. (Escherichia coli.) Was cultured for 7 days, the average value of absorbance was measured to be 0.486 when the blending ratio of zinc oxide (ZnO) was 0%, and the average value of zinc oxide ) Was 1%, the average value of the absorbance was measured to be 0.425. When the compounding ratio of zinc oxide (ZnO) was 2%, the average value of the absorbance was 0.407 and the mixing ratio of zinc oxide (ZnO) In the case of 3%, the average value of the absorbance was measured as 0.413.
도 6은 상기한 실험결과들을 실험물질별로 나타낸 그래프이며, 도 7은 상기한 실험결과들을 배양시간에 따라 나타낸 그래프이다.FIG. 6 is a graph showing the results of the experiment by the experimental materials, and FIG. 7 is a graph showing the results of the experiments according to the culturing time.
도 3 내지 도 6에 도시된 실험결과에 따르면, 실리콘과 탄소나노튜브(CNT)가 배합된 소재의 경우(징크옥사이드(ZnO) 0%) 시간이 지남에 따라 흡광도의 평균값이 급격하게 높아지며, 그에 따라 생물막의 형성이 증가됨을 알 수 있다.According to the experimental results shown in FIGS. 3 to 6, in the case of a material in which silicon and carbon nanotubes (CNT) are blended (zinc oxide (ZnO) 0%), the average value of the absorbance rapidly increases over time, The biofilm formation is increased.
한편, 실리콘과 탄소나노튜브(CNT)에 징크옥사이드(ZnO)가 1% 배합된 소재의 경우, 실리콘과 탄소나노튜브(CNT)가 배합된 소재의 경우(징크옥사이드(ZnO) 0%) 보다 흡광도의 평균값이 완만하게 증가하여, 그에 따라 생물막의 형성이 다소 억제됨을 알 수 있다.On the other hand, in the case of a material in which 1% of zinc oxide (ZnO) is blended with silicon and carbon nanotube (CNT), the absorbance is lower than that of a material in which silicon and carbon nanotube (CNT) are blended (zinc oxide (ZnO) And the formation of the biofilm is suppressed to some extent.
또한, 실리콘과 탄소나노튜브(CNT)에 징크옥사이드(ZnO)가 2% 배합된 소재의 경우, 실리콘과 탄소나노튜브(CNT)가 배합된 소재(징크옥사이드(ZnO) 0%) 및 징크옥사이드(ZnO)가 1% 배합된 소재의 경우 보다 흡광도의 평균값이 전반적으로 낮게 되어, 그에 따라 요로감염 주요 균인 E. Coli.(Escherichia Coli.)에 대한 생물막 형성 억제 효과가 명확하게 나타남을 알 수 있다.In the case of a material in which 2% of zinc oxide (ZnO) is blended with silicon and carbon nanotube (CNT), a material (zinc oxide (ZnO) 0%) and zinc oxide (carbon nanotube ZnO) was lower than that in the case of 1% mixed material, and thus the effect of inhibiting biofilm formation on E. coli (Escherichia coli.), The major strain of urinary tract infection, is clearly shown.
그리고, 실리콘과 탄소나노튜브(CNT)에 징크옥사이드(ZnO)가 5% 배합된 소재의 경우, 7일 배양시 흡광도의 평균값이 오히려 이전보다 낮아져, 장기간 사용에 따른 생물막 형성 억제 효과가 나타남을 알 수 있다.In the case of a material containing 5% of zinc oxide (ZnO) in silicon and carbon nanotube (CNT), the average value of absorbance during 7 days of culture was lower than before, .
상기한 실험결과들을 기초로 판단할 때, 실리콘과 탄소나노튜브(CNT)에 징크옥사이드(ZnO)가 약 2% 배합된 소재로 카테터 및 카테터용 폴리를 구성하는 경우, 요로감염 주요 균인 E. Coli.(Escherichia Coli.)에 대한 생물막 형성 억제 효과를 안정적으로 달성할 수 있다.Based on the above-mentioned experimental results, when a catheter and a catheter poly are composed of silicon and carbon nanotube (CNT) mixed with about 2% of zinc oxide (ZnO), E. coli (Escherichia coli.) Can be stably achieved.
도 8은 징크옥사이드(ZnO)의 배합비율이 1%인 소재에 E. Coli.(Escherichia Coli.)을 7일 동안 배양시킨 후, 실험결과를 전자현미경(Scanning Electron Microscope, SEM)으로 촬영한 것이다.8 shows the result of culturing E. coli. (Escherichia coli.) For 7 days on a material having a blending ratio of zinc oxide (ZnO) of 1% and then photographing the result of the experiment with a scanning electron microscope (SEM) .
도 8을 참조하면, 7일 배양 후에도 E. Coli. 균이 서로 뭉치거나 하여 생물막을 형성하지 않는 것을 알 수 있다.8, after 7 days of culture, E. coli. It can be seen that the microorganisms do not form a biofilm by aggregation.
또한, 아래의 표 1에 나타난 실험결과들을 참조하면, 본 발명의 일실시예에 따른 카테터 및 카테터용 폴리를 구성하는 소재는, 상기 E. Coli. 균 이외에, 포도상구균, 폐렴균, 대장균 및 녹농균에 대해서도 99.9% 이상의 살균감소율과 정균감소율을 가짐을 알 수 있다.Also, referring to the experimental results shown in Table 1 below, the material constituting the poly for the catheter and the catheter according to an embodiment of the present invention is described in E. coli. In addition to bacteria, staphylococcus, pneumococcus, Escherichia coli and Pseudomonas aeruginosa have a bactericidal reduction rate of more than 99.9% and bacteriostatic reduction rate.
한편, 상기한 바와 같이 구성된 카테터 및 카테터용 폴리는, 탄소나노튜브의 특성인 높은 열전도로 인하여 인체 삽입 과정에서 환자의 거부감을 최소화할 수 있다.On the other hand, the catheter and catheter poly configured as described above can minimize the patient's rejection feeling in the process of inserting the human body due to the high heat conduction characteristic of the carbon nanotube.
도 9는 폴리 카테터와 카테터용 폴리를 구성하는 소재의 인체 삽입시 이물감 감소 효과를 설명하기 위한 실험 결과를 나타낸 것으로, 도 9의 (a)는 탄소나노튜브가 혼합되지 않은 실리콘으로 구성된 카테터의 열전도 사진이며, 도 9의 (b)는 탄소나노튜브가 내재화된 실리콘으로 구성된 카테터의 열전도 사진이다.9 is a graph showing the results of experiments for explaining the effect of reducing the foreign body sensation upon insertion of a human body into a poly catheter and a poly catheter. FIG. 9 (a) shows the thermal conductivity 9 (b) is a thermal image of a catheter made of silicon in which carbon nanotubes are internalized.
도 9의 (a) 및 (b)를 참조하면, 탄소나노튜브가 내재화된 실리콘으로 구성된 카테터의 경우가 높은 열전도에 의해 온도의 분포가 고르게 나타남을 알 수 있다.9 (a) and 9 (b), it can be seen that in the case of a catheter made of silicon in which carbon nanotubes are internalized, the temperature distribution is uniform due to high heat conduction.
도 10은 본 발명의 일실시예에 따른 카테터용 폴리의 구성 및 동작을 설명하기 위해 도시한 것으로, 도시된 구성들 중 도 1 내지 도 9를 참조하여 설명한 것과 동일한 것에 대한 설명은 이하 생략하기로 한다.FIG. 10 is a view for explaining the configuration and operation of the catheter poly according to the embodiment of the present invention. The description of the same components as those described with reference to FIGS. 1 to 9 will be omitted hereunder do.
도 10을 참조하면, 폴리 카테터 중 방광 내에 삽입되는 일단(50)에 인접하도록 소변유입구(11)와 카테터용 폴리(200)가 카테터 본체(100)에 형성되어 있다.Referring to FIG. 10, a
카테터용 폴리(200)는 카테터의 타측에 설치되는 유체유입부(23)로부터 유체(예를 들어, 공기 또는 식염수)가 유입되면 확장되어 풍선의 형태를 가질 수 있다.The
예를 들어, 주사기 등에 유체를 미리 주입하고, 유체유입부(23)의 말단인 유입공(22)에 주사기 바늘을 끼워 놓고, 주사기를 압축시키는 방식으로 유체가 유입될 수 있다.For example, the fluid may be introduced in a manner such that a fluid is previously injected into a syringe, the syringe needle is inserted into the
카테터 본체(100)의 일단(50)에 인접하도록 형성된 카테터용 폴리(200)가 방광 내에 삽입된 상태에서, 유체유입부(23)로부터 유체통로(120, 121) 및 유체유출구(21)를 통해 유체가 카테터용 폴리(200) 내로 유입됨에 따라, 카테터용 폴리(200)가 풍선 모양으로 부풀어올라 방광목(60)에 걸쳐져 카테터를 방광 내에 고정시키게 된다.A
한편, 카테터 본체(100)의 단면의 중심부에는 소변유입구(11)와 연결된 소변통로(110)가 형성되어 있으며, 소변통로(110)의 말단에는 소변배출부(13)가 형성될다.A
예를 들어, 요도 내에서 발생하는 소변은 요도 내에 위치하는 소변유입구(11)를 통해 소변통로(110)로 유입된 후, 요도 밖에 위치하는 소변배출부(13)의 말단인 배출공(12)을 통해 배출될 수 있다.For example, the urine generated in the urethra flows into the
도 11은 본 발명에 따른 카테터용 폴리의 구성에 대한 일실시예를 보다 상세히 도시한 것으로, 도 1 및 도 11에 도시된 폴리 카테터 중 카테터용 폴리(200)가 접합되어 있는 부분을 확대하여 나타낸 것이다.Fig. 11 shows one embodiment of the configuration of the
도 11을 참조하면, 카테터용 폴리(200)의 양 끝단에 카테터 본체(100)와 접합되기 위한 접합면들(210, 211)이 형성되어 있으며, 카테터 본체(100)에 카테터용 폴리(200)를 형성시키는 과정에서 접착제를 이용하여 접합면들(210, 211)을 카테터 본체(100)에 접착시킬 수 있다.11, bonding surfaces 210 and 211 are formed at both ends of the
이하, 본 발명의 일실시예에 따른 폴리 카테터 제조 방법 및 카테터용 폴리의 형성 방법에 대해 도 11을 참조하여 설명하기로 한다.Hereinafter, a method for manufacturing a poly catheter and a method for forming a poly for a catheter according to an embodiment of the present invention will be described with reference to FIG.
먼저, 상기한 바와 같이 탄소나노튜브와 징크옥사이드(ZnO)가 결합된 탄소나노튜브 중합체(CNT Polymer)가 실리콘에 배합된 소재를 튜브형 관체로 압출하여 카테터 본체(100)를 생성한다.First, as described above, a carbon nanotube polymer (CNT polymer) in which carbon nanotubes and zinc oxide (ZnO) are combined is extruded into a tubular tube to form a
또한, 탄소나노튜브와 징크옥사이드(ZnO)가 결합된 탄소나노튜브 중합체(CNT Polymer)가 실리콘에 배합된 소재를 튜브형 관체로 압출 또는 사출하여 카테터용 폴리(200)를 생성한다.In addition, the carbon nanotube and zinc oxide (ZnO) -containing carbon nanotube polymer (CNT polymer) are extruded or injected into a tubular tube to form a poly (200) for a catheter.
여기서, 카테터 본체(100)를 생성하기 위한 소재의 탄소나노튜브, 징크옥사이드(ZnO) 또는 탄소나노튜브 중합체의 배합비율과, 카테터용 폴리(200)를 생성하기 위한 소재의 탄소나노튜브, 징크옥사이드(ZnO) 또는 탄소나노튜브 중합체의 배합비율이 상이할 수 있다.Here, the blending ratio of the carbon nanotube, zinc oxide (ZnO), or carbon nanotube polymer as a raw material for producing the
예를 들어, 카테터용 폴리(200)가 외부로부터 유입되는 유체에 의해 팽창되면 표면적이 증가하게 되고, 그에 따라 단위면적당 탄소나노튜브 중합체의 분포가 감소하여 항균력이 떨어질 수 있다.For example, when the poly (200) for a catheter is inflated by a fluid introduced from the outside, the surface area is increased, and thus the distribution of the carbon nanotube polymer per unit area is decreased, and the antibacterial ability may be lowered.
위와 같이, 카테터용 폴리(200)가 팽창하여 항균력이 떨어지면, 카테터 본체(100)에서는 생물막의 형성이 억제되더라도, 향균력이 떨어진 카테터용 폴리(200)에서 생물막이 형성되어 세균 감염의 원인이 될 수 있다.As described above, if the poly (200) for catheter is inflated and the antibacterial power is lowered, even if formation of biofilm is inhibited in the
본 발명의 일실시예에 따르면, 카테터용 폴리(200)의 탄소나노튜브 중합체 배합 비율은 카테터 본체(100)의 탄소나노튜브 중합체 배합 비율보다 높은 것이 바람직하며, 그에 따라 카테터용 폴리(200)가 팽창하더라도 단위면적당 탄소나노튜브 중합체의 분포가 카테터 본체(100)와 동일 또는 유사한 범위를 가져 폴리 카테터의 위치에 관계없이 균일한 향균력을 유지할 수 있다.According to one embodiment of the present invention, the ratio of the carbon nanotube polymer of the
카테터용 폴리(200)가 팽창하여 표면적이 약 4배 내지 6배 증가한다고 가정하면, 상기한 바와 같이 카테터 본체(100)가 실리콘 100 중량부 대비 탄소나노튜브 중합체 1.0 내지 2.2 중량부로 배합된 소재로 구성될 때, 카테터용 폴리(200)는 표면적의 증가량에 비례하여 실리콘 100 중량부 대비 탄소나노튜브 중합체 4.0 내지 13.2 중량부로 배합된 소재로 구성될 수 있다.Assuming that the
한편, 카테터용 폴리(200)의 경우, 상기 화학식 1에서 n과 m의 값이, 카테터용 폴리(200)의 팽창 시 표면적 증가량에 비례하도록 증가되는 것이 바람직할 수 있다.On the other hand, in the case of the poly (200) for a catheter, it may be desirable that the values of n and m in the formula (1) increase in proportion to the increase in surface area upon expansion of the catheter poly (200).
카테터 본체(100)와 카테터용 폴리(200)가 준비된 후에, 카테터용 폴리(200)의 접합면들(210, 211)을 카테터 본체(100)에 접착시키고, 카테터 본체(100)의 일단에 소변유입구(11)가 구비된 팁(tip)을 형성시켜 폴리 카테터를 제조할 수 있다.After the
본 발명의 일실시예에 따른 폴리 카테터 제조 방법 및 카테터용 폴리의 형성 방법은 상기에서 설명한 단계들 이외에 건조 단계 등과 같은 추가적인 단계들을 더 포함할 수 있으며, 이하에서 설명할 추가적인 구성을 위한 별도의 단계들이 더 추가될 수도 있다.The method of manufacturing a poly catheter according to an embodiment of the present invention and the method of forming a poly for a catheter may further include additional steps such as a drying step in addition to the steps described above and a separate step May be added.
도 12 내지 도 16은 도 11에 도시된 카테터용 폴리의 단면 구성에 대한 예들을 단면도들로 도시한 것이다.Figs. 12-16 show cross-sectional views of examples of the cross-sectional configuration of the catheter poly shown in Fig.
도 12는 도 11에 도시된 C-C선에 따른 횡단면도를 나타낸 것으로, 소변통로(110)와 유체통로(120, 121)가 구비된 카테터 본체(100)를 카테터용 폴리(200)가 감싸고 있는 구조를 가질 수 있다.12 is a cross-sectional view taken along the CC line shown in Fig. 11, showing a structure in which the
여기서, 상기한 바와 같이 카테터용 폴리(200)가 카테터 본체(100) 보다 탄소나노튜브 중합체의 배합비율이 높은 소재를 사용하여 제조된 경우, 도 12에 도시된 카테터 본체(100)와 카테터용 폴리(200) 사이에 전위차가 없어질 때까지 전류가 흐를 수 있다.Here, when the
이는, 징크옥사이드(ZnO)가 높은 전위를 만들어, 서로 접하는 부분에서 카테터용 폴리(200)로부터 카테터 본체(100)로 전류가 흐르게 되는 것으로서, 카테터 본체(100)와 카테터용 폴리(200) 사이에 전위차가 없어지게 되면 카테터용 폴리(200)가 팽창하는 경우 정전용량이 낮아져 항균력이 감소하게 될 수 있다.This is because zinc oxide (ZnO) creates a high dislocation and current flows from the
위와 같이 카테터용 폴리(200)로부터 카테터 본체(100)로 전류가 흐르는 것을 방지하기 위하여, 도 13에 도시된 바와 같이 카테터 본체(100)와 카테터용 폴리(200) 사이에 절연층(900)이 형성될 수 있다.An insulating
상기 절연층(900)은 절연층은 공기층 또는 멸균 가스층(예를 들어, EO 가스) 등의 기체로 구성되거나, 또는 고농도의 탄소나노튜브 코팅층으로 구성될 수 있다.The insulating
상기 절연층(900)에 의해 카테터용 폴리(200)로부터 카테터 본체(100)로 전류가 흐르는 것이 방지됨으로써, 탄소나노튜브 중합체 배합 비율의 차이에 따른 전위차가 유지될 수 있으며, 그로 인해 카테터용 폴리(200)의 팽창 시에도 카테터 본체(100)와 동일 또는 유사한 범위에서 항균력이 유지될 수 있다.By preventing the electric current from flowing from the
도 13에 도시된 바와 같이 카테터 본체(100)와 카테터용 폴리(200) 사이에 절연층(900)이 형성되도록, 카테터용 폴리(200)의 형성 과정에서 카테터 본체(100)와 카테터용 폴리(200) 사이에 공기층이 주입되도록 하거나 또는 카테터 본체(100)의 외면에 EO 가스 처리 또는 고농도의 탄소나노튜브 코팅 처리를 하는 단계가 추가될 수 있다.The
도 14는 도 11에 도시된 D-D선에 따른 횡단면도를 나타낸 것으로, 카테터용 폴리(200)가 카테터 본체(100)에 접합되는 접합면(210) 부분의 단면 구조를 도시한 것이다.FIG. 14 is a cross-sectional view taken along the line D-D shown in FIG. 11, showing a cross-sectional structure of a portion of the
도 14를 참조하면, 카테터용 폴리(200)가 카테터 본체(100)에 접합되는 접합면(210)에 절연막(1000)이 형성될 수 있다.Referring to FIG. 14, an insulating
이는, 카테터용 폴리(200)의 접합면(210)을 카테터 본체(100)에 접합시키기 위해 사용되는 접착제가 도전성을 가지는 경우, 탄소나노튜브 중합체 배합 비율의 차이에 따른 전위차에 의해 접합면(210)을 통해 카테터용 폴리(200)로부터 카테터 본체(100)로 전류가 흐를 수 있으므로, 절연막(1000)을 이용하여 전류의 흐름을 차단하기 위한 것이다.This is because when the adhesive used to bond the
예를 들어, 절연막(1000)은 탄소나노튜브 절연막으로서, 고농도의 탄소나노튜브를 접합면(210)에 코팅하여 형성될 수 있으며, 카테터용 폴리(200)가 팽창되는 경우 표면적의 증가로 인해 자연스럽게 해체될 수도 있다.For example, the insulating
도 14에 도시된 바와 같은 절연막(1000)은 심혈관용 카테터에 대해서도 적용되는 것이 바람직하다.The insulating
도 15는 도 11에 도시된 E-E선에 따른 횡단면도를 나타낸 것으로, 유체유출구(21)가 형성된 부분의 단면 구조를 도시한 것이다.FIG. 15 is a cross-sectional view taken along the line E-E shown in FIG. 11, showing a cross-sectional structure of a portion where the
도 15를 참조하면, 카테터 본체(100)에 형성된 두 개의 유체통로들(120, 121)과 연통하도록 각각에 대응되는 위치에 유체 유출구들(21a, 21b)이 형성되어 있을 수 있다.Referring to FIG. 15,
그에 따라, 외부로부터 유입된 유체는 카테터 본체(100)의 유체통로들(120, 121)을 통해 이동한 후, 유체 유출구들(21a, 21b)을 통해 카테터용 폴리(200)로 유출되어 카테터용 폴리(200)를 유체의 압력에 의해 팽창시킨다.The fluid introduced from the outside flows through the
도 16은 카테터용 폴리(200)가 팽창된 상태를 나타내기 위해 도시한 단면도이다.16 is a cross-sectional view illustrating the expanded state of the poly (200) for catheter.
도 16을 참조하면, 상기한 바와 같이 팽창 시 표면적 증가비율에 비례하도록 카테터용 폴리(200)의 탄소나노튜브 중합체 배합 비율을 카테터 본체(100)의 탄소나노튜브 중합체 배합 비율보다 크게 함으로써, 팽창 시에도 카테터용 폴리(200)의 정전용량이 카테터 본체(100)와 동일 또는 유사하게 유지되어 폴리 카테터의 향균력이 균일하도록 할 수 있다.Referring to FIG. 16, by increasing the compounding ratio of the carbon nanotube polymer of the poly (200) for catheter to that of the
도면 상에 별도로 도시하지는 않았으나, 본 발명에 따른 카테터 또는 카테터용 폴리의 종류나 기능에 따라 상기 제조방법에 따른 단계들은 추가되거나 축소될 수 있다.Although not shown separately in the drawing, steps according to the manufacturing method may be added or reduced depending on the type or function of the catheter or catheter poly according to the present invention.
이상에서는 본 발명의 바람직한 실시예에 대하여 도시하고 설명하였지만, 본 발명은 상술한 특정의 실시예에 한정되지 아니하며, 청구범위에서 청구하는 본 발명의 요지를 벗어남이 없이 당해 발명이 속하는 기술분야에서 통상의 지식을 가진 자에 의해 다양한 변형 실시가 가능한 것은 물론이고, 이러한 변형 실시들은 본 발명의 기술적 사상이나 전망으로부터 개별적으로 이해되어져서는 안 될 것이다.While the present invention has been particularly shown and described with reference to exemplary embodiments thereof, it is clearly understood that the same is by way of illustration and example only and is not to be construed as limiting the scope of the invention as defined by the appended claims. It will be understood by those skilled in the art that various changes in form and details may be made therein without departing from the spirit and scope of the present invention.
Claims (4)
분산된 탄소나노튜브와 징크옥사이드(ZnO)를 실리콘에 복합 처방하여 탄소나노튜브 중합체(CNT Polymer)가 실리콘에 배합된 소재를 구성하는 폴리 카테터용 조성물 제조 방법.A method of making a composition for use in a poly catheter,
A method for manufacturing a composition for a poly catheter, comprising the step of compounding dispersed carbon nanotubes and zinc oxide (ZnO) with silicon to form a material in which carbon nanotube polymer (CNT polymer) is blended with silicon.
카테터용 폴리에 사용되는 경우, 상기 탄소나노튜브 중합체는 실리콘 100 중량부 대비 4.0 내지 13.2 중량부로 배합되는 폴리 카테터용 조성물 제조 방법.The method according to claim 1,
Wherein the carbon nanotube polymer is blended in an amount of 4.0 to 13.2 parts by weight based on 100 parts by weight of silicone when used in a catheter poly.
카테터 본체에 사용되는 경우, 상기 탄소나노튜브 중합체는 실리콘 100 중량부 대비 1.0 내지 2.2 중량부로 배합되는 폴리 카테터용 조성물 제조 방법.The method according to claim 1,
Wherein the carbon nanotube polymer is used in an amount of 1.0 to 2.2 parts by weight based on 100 parts by weight of silicone when used in a catheter body.
다중벽 탄소나노튜브(Multi-Walled Carbon Nanotube, MWNT)인 폴리 카테터용 조성물.The method of claim 1, wherein the carbon nanotubes
A composition for a poly catheter, the multi-walled carbon nanotube (MWNT).
Priority Applications (4)
Application Number | Priority Date | Filing Date | Title |
---|---|---|---|
US16/068,173 US20190015560A1 (en) | 2016-02-11 | 2017-02-02 | Composition for catheter and production method therefor |
CN201780006276.2A CN108463489B (en) | 2016-02-11 | 2017-02-02 | Composition for catheter and method for producing the same |
PCT/KR2017/001138 WO2017138716A1 (en) | 2016-02-11 | 2017-02-02 | Composition for catheter and production method therefor |
US16/991,435 US20200368399A1 (en) | 2016-02-11 | 2020-08-12 | Catheter, composition for catheter, production method therefor |
Applications Claiming Priority (2)
Application Number | Priority Date | Filing Date | Title |
---|---|---|---|
KR20160015841 | 2016-02-11 | ||
KR1020160015841 | 2016-02-11 |
Publications (1)
Publication Number | Publication Date |
---|---|
KR20170094758A true KR20170094758A (en) | 2017-08-21 |
Family
ID=59757425
Family Applications (1)
Application Number | Title | Priority Date | Filing Date |
---|---|---|---|
KR1020170013679A KR20170094758A (en) | 2016-02-11 | 2017-01-31 | Method of Manufacturing Composition for Catheter |
Country Status (1)
Country | Link |
---|---|
KR (1) | KR20170094758A (en) |
-
2017
- 2017-01-31 KR KR1020170013679A patent/KR20170094758A/en not_active Application Discontinuation
Similar Documents
Publication | Publication Date | Title |
---|---|---|
Oliveira et al. | Staphylococcus aureus and Staphylococcus epidermidis infections on implants | |
KR102448392B1 (en) | catheter with varying characteristics | |
Lawrence et al. | Materials for urinary catheters: a review of their history and development in the UK | |
AU2017397418B2 (en) | Catheter system for continuous irrigation | |
Holban et al. | Inorganic nanoarchitectonics designed for drug delivery and anti-infective surfaces | |
WO2016087926A1 (en) | Bladder catheter for the minimally invasive discharge of urine | |
DE60329882D1 (en) | Medical hose and extrusion nozzle for its manufacture | |
ATE249850T1 (en) | ADJUSTABLE MEMBRANE BALLOON MADE OF PEEK MATERIAL | |
KR101900739B1 (en) | Foley Catheter | |
Jones et al. | Physicochemical characterization of hexetidine-impregnated endotracheal tube poly (vinyl chloride) and resistance to adherence of respiratory bacterial pathogens | |
EP2099370B1 (en) | Tube for enteral nutrition | |
KR20170094755A (en) | Method for manufacturing Foley Catheter | |
KR20170094753A (en) | Method for forming Foley for Catheter | |
CN108463489B (en) | Composition for catheter and method for producing the same | |
US20180272039A1 (en) | Foley catheter and method for manufacturing same | |
KR20180103803A (en) | Foley Catheter | |
KR20170094758A (en) | Method of Manufacturing Composition for Catheter | |
US20200368399A1 (en) | Catheter, composition for catheter, production method therefor | |
KR20170094757A (en) | Composition for Catheter Foley | |
KR20170094756A (en) | Composition for Foley Catheter | |
KR20170094752A (en) | Foley for Catheter | |
US6165158A (en) | Lubricious catheter shaft | |
Jiang et al. | Preparation of silicon‐modified antimicrobial polyethylene endotracheal tubes | |
KR20160124962A (en) | Fabrication method of antibacterial CNT Catheter with biofilm forming prevention | |
CN202185046U (en) | Disposable bacteriostatic ultra-smooth medical indwelling catheter |
Legal Events
Date | Code | Title | Description |
---|---|---|---|
A201 | Request for examination | ||
E902 | Notification of reason for refusal | ||
E601 | Decision to refuse application |