TWI409448B - Process for measurements of concentration of dissolved organic nitrogen (don) with low concentration in water - Google Patents
Process for measurements of concentration of dissolved organic nitrogen (don) with low concentration in water Download PDFInfo
- Publication number
- TWI409448B TWI409448B TW99126923A TW99126923A TWI409448B TW I409448 B TWI409448 B TW I409448B TW 99126923 A TW99126923 A TW 99126923A TW 99126923 A TW99126923 A TW 99126923A TW I409448 B TWI409448 B TW I409448B
- Authority
- TW
- Taiwan
- Prior art keywords
- concentration
- water
- water sample
- nitrogen
- don
- Prior art date
Links
Abstract
Description
本發明係有關於一種測定水中低濃度溶解性有機氮濃度的方法,尤其是指一種於環境保護、水處理領域,涉及水中溶解性有機氮〔DON〕測定的有效方法。The invention relates to a method for determining the concentration of dissolved organic nitrogen in water at low concentration, in particular to an effective method for determining the dissolved organic nitrogen [DON] in water in the field of environmental protection and water treatment.
按,當前環境水體普遍受微污染情況嚴重,原水中常見有機物含量超過原水水質標準、含氮化合物濃度高、水體有異味、色度偏高、及藻類大量繁殖等問題。鑒於水中廣泛存在的溶解性有機物〔DOM〕極易在消毒過程中與氯發生鹵化反應,而形成三鹵甲烷、鹵乙酸、鹵乙腈、呋喃酮〔MX〕等氯化消毒副產物,因此對微污染原水中有機物的消除與控制,一直以來是本領域持續關注的焦點。溶解性有機氮〔DON〕在溶解性有機物〔DOM〕中佔0.5-10%左右,主要包含NH類、氨基類、腈類、嘌呤、嘧啶、硝基化合物等。由於DON絕大部分物質本身對人體具有直接或間接致癌或致突變性,同時這些化合物的存在對飲用水處理會產生眾多不利的影響,例如消耗氧化劑和 消毒劑、產生更多的鹵乙酸和三鹵甲烷、造成薄膜積垢等等,因此對DON的認識和控制已逐步成為水領域研究中的新興重點,尤其是對DON的強致癌性氯化副產物-鹵化硝基甲烷〔HNMs〕與二甲基亞硝胺〔包含NDMA〕的產生與消除,也已成為國際近來關注的重要問題。作為HNMs與二甲基亞硝胺〔包含NDMA〕的前驅物質,DON更是引起人們的關注。According to the current situation, the current environmental water body is generally subject to micro-pollution. The common organic matter content in the raw water exceeds the raw water quality standard, the concentration of nitrogen-containing compounds is high, the water body has odor, the chroma is high, and the algae multiply. In view of the widespread presence of dissolved organic matter (DOM) in water, it is easy to undergo halogenation reaction with chlorine during the disinfection process, and forms chlorination by-products such as trihalomethane, haloacetic acid, haloacetonitrile, furanone [MX], etc. The elimination and control of organic matter in polluted raw water has been the focus of continuous attention in this field. The dissolved organic nitrogen [DON] accounts for about 0.5-10% of the dissolved organic matter [DOM], and mainly contains NHs, amino acids, nitriles, purines, pyrimidines, nitro compounds, and the like. Since most of DON's substances are directly or indirectly carcinogenic or mutagenic to humans, the presence of these compounds can have many adverse effects on drinking water treatment, such as the consumption of oxidants and Disinfectants, produce more haloacetic acid and trihalomethane, cause film fouling, etc., so the understanding and control of DON has gradually become an emerging focus in the field of water research, especially the strong carcinogenic chlorination of DON The production and elimination of the product-halogenated nitromethane [HNMs] and dimethylnitrosamine (including NDMA) has also become an important issue of international concern. As a precursor of HNMs and dimethyl nitrosamines (including NDMA), DON has attracted people's attention.
目前DON的測定尚沒有直接的方法,主要通過以計算總溶解性氮〔TDN〕和無機氮〔DIN〕(包含硝酸鹽氮、亞硝酸鹽氮和氨氮)的差值來求得。由於地表水中DON含量一般比DIN小一個數量級以上,因此DON的分析誤差往往被擴大,例如TN為1mg/L±0.1、DIN為0.9mg/L±0.1,此時得到的DON值將是0.1mg/L±0.2,誤差值將達到200%。在實際水樣DIN較高的情況下〔例如黃浦江水TDN為3-6mg/L,NO3 - 為2-4mg/L,NH4 + -N為0.1-2mg/L;DIN/TDN>80%〕,由於分析儀器誤差和測試方法精度不夠(DIN的精度為0.1mg/L),試驗發現按照標準方法測定各水質指標,仍然有20-30%左右的水樣因DIN濃度值大於測定的TDN值,導致DON濃度計算後為負值,這與國外研究情況一致,因此嚴重地影響DON及其含氮消毒副產物〔N-DBPs〕的研究工作。At present, there is no direct method for the determination of DON, which is mainly obtained by calculating the difference between total dissolved nitrogen [TDN] and inorganic nitrogen [DIN] (including nitrate nitrogen, nitrite nitrogen and ammonia nitrogen). Since the DON content in surface water is generally more than an order of magnitude smaller than DIN, the analysis error of DON is often expanded, for example, TN is 1 mg/L ± 0.1, DIN is 0.9 mg/L ± 0.1, and DON is obtained at this time. The value will be 0.1 mg/L ± 0.2 and the error value will reach 200%. In the case where the actual water sample DIN is high (for example, the TDN of Huangpu River is 3-6 mg/L, NO 3 - is 2-4 mg/L, NH 4 + -N is 0.1-2 mg/L; DIN/TDN >80%], due to the error of the analytical instrument and the accuracy of the test method (the accuracy of DIN is 0.1mg/L), the test found that the water quality index is determined according to the standard method, and there are still about 20-30% of the water sample due to the DIN concentration value. Greater than the measured TDN value, resulting in a negative value after the DON concentration calculation, which is consistent with foreign research, thus seriously affecting the research work of DON and its nitrogen-containing disinfection by-products [N-DBPs].
針對此一難題,有美國研究員提出先把DIN去除/分離後,再進行總氮的分析,以提高水中DON量測的準確度, 其主要做法包括:(1)採用一定分子量的透析膜讓水中大部分DIN透出,而將DON保留下來、(2)採用催化還原的方法,將硝酸鹽進行還原而去除。由於採用第二種方法可能使得DON吸附在催化劑上,且分離效果受還原反應條件影響較大,因此實際應用在DON分析上的可行性較低;而第一種方法,目前國際上已有採用100Dalton的透析膜來去除DIN,可有效應用於DIN/TDN>0.6的水樣DON分析,但該方法的裝置、透析管和透析膜成本非常高、所需的超純水量大(至少需144公升)、分析時間長(至少需24小時以上),故在實際應用上仍具有很大的局限性。In response to this problem, some American researchers have proposed to remove and separate DIN before performing total nitrogen analysis to improve the accuracy of DON measurement in water. The main practices include: (1) using a dialysis membrane of a certain molecular weight to allow most of the DIN in the water to pass out while retaining DON, and (2) using a catalytic reduction method to reduce and remove the nitrate. Since the second method may cause DON to adsorb on the catalyst, and the separation effect is greatly affected by the reduction reaction conditions, the practical application is less feasible in the DON analysis; and the first method has been adopted internationally. 100Dalton dialysis membrane to remove DIN, can be effectively applied to DON analysis of water sample with DIN/TDN>0.6, but the device, dialysis tube and dialysis membrane of this method are very expensive and require a large amount of ultrapure water (at least 144 liters), long analysis time (at least 24 hours or more), so it still has great limitations in practical applications.
有鑑於此,本發明人秉持多年該相關行業之豐富設計開發及實際製作經驗,特別提供一種測定水中低濃度溶解性有機氮濃度的方法,來達成水中溶解性有機氮〔DON〕測定的有效方法,以期具有更佳實用價值性之目的者。In view of this, the inventors have long been rich in design and development and practical production experience in the relevant industries, and in particular, provide an effective method for determining the concentration of dissolved organic nitrogen in water to achieve the determination of dissolved organic nitrogen [DON] in water. For the purpose of having a better practical value.
為達到以上目的,本發明所採用的解決方案是:應用奈米薄膜分離技術來提昇溶解性有機氮〔DON〕濃度量測的準確度。奈米薄膜過濾是介於逆滲透和超過濾之間的一種壓力驅動膜分離過程,對高價離子和有機物具有較高的去除率,而對於一價離子卻具有很高的通過率,此正好符合DON量測問題的解決關鍵,此為本發明方法對DON測定準確度提昇的主要目的者。In order to achieve the above object, the solution adopted by the present invention is to apply nano film separation technology to improve the accuracy of the measurement of dissolved organic nitrogen [DON] concentration. Nanofiltration is a pressure-driven membrane separation process between reverse osmosis and ultrafiltration. It has a high removal rate for high-valent ions and organic matter, and a high throughput for monovalent ions, which is in line with The key to solving the DON measurement problem is the main purpose of the method for improving the accuracy of DON measurement.
本發明為一種準確測定溶解性有機氮濃度的方法, 利用奈米膜預處理以提昇水中溶解性有機氮量測的準確度。在預處理之前的準備過程中,水樣用孔徑大於等於0.22μm而小於等於0.45μm的微濾膜過濾以除去顆粒性雜質,取過濾後的水樣調節其pH值範圍於6.0-9.0,以截留分子量為150-500Dalton的選擇性奈米薄膜,利用垂直〔dead-end〕過濾或掃流〔cross-flow〕過濾方式,將水樣中的DON濃縮6-15倍;最後水樣中的DON含量,可由總氮減去其他測定之無機氮含量總和(包含硝酸鹽氮、亞硝酸鹽氮和氨氮)後,除以實際濃縮倍數而獲得;藉此,提高了DON測試的準確度和精確度,解決了直接水樣測定法中DON經常出現負值、濃度標準偏差大等問題;且整個測定過程簡單有效、易於操作、成本低,可廣泛應用於地表和地下水監測,以及大、中和小型淨水場與污水廠的監測與分析。The invention is a method for accurately determining the concentration of dissolved organic nitrogen, Nanofilm pretreatment is used to increase the accuracy of the measurement of dissolved organic nitrogen in water. In the preparation process before pretreatment, the water sample is filtered with a microfiltration membrane having a pore diameter of 0.22 μm or more and 0.45 μm or less to remove particulate impurities, and the filtered water sample is adjusted to have a pH range of 6.0. - 9.0, a selective nano film with a molecular weight cut off of 150-500 Dalton, using a vertical (dead-end) filtration or cross-flow filtration method to concentrate DON in the water sample 6-15 times; The DON content in the final water sample can be obtained by subtracting the sum of other measured inorganic nitrogen contents (including nitrate nitrogen, nitrite nitrogen and ammonia nitrogen) from the total nitrogen, and dividing by the actual concentration factor; thereby, the DON test is improved. The accuracy and accuracy of the direct water sample method solve the problems of DON often negative values and large standard deviation of the concentration; and the whole measurement process is simple and effective, easy to operate, low cost, and can be widely applied to surface and groundwater monitoring. And monitoring and analysis of large, medium and small water purification plants and sewage plants.
具體來說,本發明進一步包含:環境水樣在奈米過濾前的前處理、奈米過濾預處理過程中參數的控制與操作方式、相關水質分析〔如:總氮、硝酸鹽氮、亞硝酸鹽氮、氨氮〕、及最後DON的濃度計算。在奈米過濾前的準備中,環境樣品先用孔徑為0.22或0.45μm的醋酸纖維〔cellulose acetate〕膜或聚醚碸〔polyestersulfone)的微濾膜過濾以除去顆粒性雜質〔這兩種膜功能與其他膜相比具有孔隙率低、對蛋白質吸附率低、對大分子截留效果好等特點,而且不含氮元素,不會因為膜釋放出氮而對後續水樣測定 DON產生影響〕;之後取過濾後水樣約300-500mL,使用鹼(例如氫氧化鈉〔NaOH〕)及酸(例如鹽酸〔HCl〕)調節pH值範圍為6-9後,採用截留分子量為150-500Dalton的選擇性奈米薄膜進行垂直〔dead-end〕或掃流〔cross-flow〕方式過濾,使水樣中DON濃縮約6-15倍。水質分析分別採用總氮分析儀、離子層析儀和分光光度計測定濃縮液中總氮、硝酸鹽氮、亞硝酸鹽氮和氨氮的含量;水樣DON含量通過總氮減去其他測定無機氮含量總和後,除以實際濃縮倍數而求得。Specifically, the present invention further comprises: pretreatment of the environmental water sample before the nanofiltration, parameter control and operation mode during the nanofiltration pretreatment process, and related water quality analysis (eg, total nitrogen, nitrate nitrogen, nitrous acid) The concentration of salt nitrogen, ammonia nitrogen, and finally DON was calculated. In the preparation before the nanofiltration, the environmental sample is first filtered with a microfiltration membrane of a cellulose acetate membrane or a polyethersulfone having a pore size of 0.22 or 0.45 μm to remove particulate impurities. Compared with other membranes, the membrane function has the characteristics of low porosity, low protein adsorption rate and good macromolecular retention, and does not contain nitrogen. It does not determine the subsequent water samples due to the release of nitrogen from the membrane. DON has an effect]; after taking the filtered water sample about 300-500mL, using a base (such as sodium hydroxide [NaOH]) and an acid (such as hydrochloric acid [HCl]) to adjust the pH range of 6-9, the molecular weight cutoff is The 150-500 Dalton selective nanofilm is subjected to a dead-end or cross-flow filtration to concentrate the DON in the water sample by about 6-15 times. Water quality analysis used total nitrogen analyzer, ion chromatograph and spectrophotometer to determine the content of total nitrogen, nitrate nitrogen, nitrite nitrogen and ammonia nitrogen in the concentrate; the water sample DON content was determined by total nitrogen minus other inorganic nitrogen After the sum of the contents, it is obtained by dividing by the actual concentration factor.
其中pH值的選取進一步經由以下試驗方式來決定:係採用自來水廠清水進行試驗,使用HCl、NaOH調節通過0.22或0.45μm的微濾膜後水樣之pH值,應用本發明方法,分析不同pH條件下相同水樣DON的測定值,從而探求pH對選擇性奈米薄膜的影響,得出DON測定值最大時的最佳pH條件。The selection of the pH value is further determined by the following test method: the water is tested by using the water of the water plant, and the pH value of the water sample after passing through the microfiltration membrane of 0.22 or 0.45 μm is adjusted by using HCl and NaOH, and the method of the invention is applied. The measured values of DON in the same water sample under different pH conditions were analyzed to explore the effect of pH on the selective nanofilm, and the optimum pH condition when the DON value was the largest was obtained.
其中合適的濃縮倍數選取進一步以下試驗方案確定:係採用不同體積的自來水廠清水進行試驗,使用HCl、NaOH調整通過0.22或0.45μm的微濾膜後水樣之pH值至6.5-7.5,應用本發明提出的方法,控制經奈米薄膜預處理後濃縮液的體積均為約50mL,分析不同濃縮倍數下所量測該水樣DON的濃度,從而評定濃縮倍數對DON量測的影響,以求得DON測定值最大的最佳濃縮倍數。The suitable concentration factor is further determined by the following test scheme: the water is tested by using different volumes of water in the waterworks, and the pH of the water sample after passing through the 0.22 or 0.45 μm microfiltration membrane is adjusted to 6.5 using HCl and NaOH. -7.5, applying the method proposed by the invention, controlling the volume of the concentrated liquid after pretreatment by nano film is about 50 mL, analyzing the concentration of the water sample DON under different concentration multiples, thereby evaluating the concentration multiple to DON The effect of the measurement is taken to obtain the best concentration factor for the largest DON measurement.
其中截留分子量為150-500Dalton的選擇性奈米薄膜類型選擇,進一步以下列試驗方案確定:係採用相同體積的自來水廠清水進行試驗,使用HCl、NaOH調整通過0.22或0.45μm的微濾膜後水樣之pH值至6.5-7.5,並控制濃縮倍數為6倍左右,分析不同類型奈米薄膜對DON測定值的影響,得出DON測定值最大、且重複試驗後標準偏差最小的奈米薄膜種類。The selection of selective nanofilm type with molecular weight cutoff of 150-500 Dalton is further determined by the following test scheme: using the same volume of water from the waterworks for testing, using HCl, NaOH to adjust the microfiltration through 0.22 or 0.45 μm. The pH value of the post-membrane water sample was 6.5-7.5, and the concentration ratio was controlled to be about 6 times. The influence of different types of nano-films on the measured value of DON was analyzed, and the maximum value of DON was obtained, and the standard after repeated test was obtained. The type of nano film with the smallest deviation.
(1)‧‧‧水樣(1)‧‧‧ water samples
(2)‧‧‧奈米薄膜(2)‧‧‧Nano film
(3)‧‧‧過濾設備(3) ‧‧‧Filter equipment
(31)‧‧‧過濾杯(31)‧‧‧Filter Cup
(32)‧‧‧磁石力攪拌機(32)‧‧‧Magnetic mixer
(4)‧‧‧容器(4) ‧ ‧ container
(5)‧‧‧氮氣(5) ‧‧‧nitrogen
第一圖:本發明奈米薄膜預處理垂直過濾方式的裝置示意圖First: Schematic diagram of the apparatus for pretreating the vertical membrane of the nano film of the present invention
第二圖:本發明奈米薄膜預處理掃流過濾方式的流程示意圖The second figure: the flow chart of the pretreatment sweep filtration method of the nano film of the invention
第三圖:本發明實際操作流程圖Third figure: actual operation flow chart of the present invention
第四圖:本發明不同奈米薄膜類型對測定結果的影響比較圖Figure 4: Comparison of the effects of different nanofilm types on the determination results of the present invention
第五圖:本發明不同pH值測定條件下對量測結果的影響比較圖Figure 5: Comparison of the effects of different pH values on the measurement results of the present invention
第六圖:本發明不同濃縮倍數量測結果的影響比較圖Figure 6: Comparison of the effects of different concentration times on the measurement results of the present invention
第七圖:本發明直接量測及應用本預處理方法對微污染黃浦江原水及清水中DON量測結果比較圖Figure 7: Comparison of direct measurement and application of the pretreatment method to the DN measurement results of micro-polluted Huangpu River raw water and clean water
為令本發明之技術手段、發明目的及達成功效有更完整及清楚的揭露,茲於下詳細說明之,並請一併參閱揭示之圖式及元件符號。For a more complete and clear disclosure of the technical means, the object of the invention and the effect of the present invention, the following detailed description, and the accompanying drawings,
首先,請參閱第一~三圖所示,為本發明一種測定溶解性有機氮濃度的方法之過濾裝置與流程示意圖,其具體步驟:First, please refer to the first to third figures, which is a schematic diagram of a filtering device and a flow chart of a method for determining the concentration of dissolved organic nitrogen according to the present invention, and the specific steps thereof are as follows:
(S1):為預處理之前的準備過程,係將待測水樣(1)先通過0.22或0.45μm的醋酸纖維〔cellulose acetate〕或聚醚碸〔polyestersulfone〕微濾膜進行過濾,以去除水中懸浮物與顆粒雜質〔如不能即時進行測定,需將待測水樣放入4℃冰箱保存〕。(S1): For the preparation process before pretreatment, the water sample to be tested (1) is first filtered through a 0.22 or 0.45 μm cellulose acetate or polyestersulfone microfiltration membrane. In order to remove suspended solids and particulate impurities in water [if it is not possible to measure immediately, the water sample to be tested should be stored in a refrigerator at 4 ° C].
(S2):再使用氫氧化鈉〔NaOH〕和氯化氫〔HCl〕調節待測水樣的pH值到6.0-9.0。(S2): The pH of the water sample to be tested is adjusted to 6.0-9.0 by using sodium hydroxide [NaOH] and hydrogen chloride [HCl].
(S3):將用超純水浸泡過〔不低於24小時〕的選擇性奈米薄膜(2)放置到過濾設備(3)中〔此處使用垂直過濾的過濾杯(31)〕,出水口放置一潔淨容器(4)收集濾液;過濾設備(3)應密封以防止樣品水份的揮發。(S3): Place the selective nano film (2) soaked in ultrapure water [not less than 24 hours] into the filtering device (3) (using a vertically filtered filter cup (31)] The filtrate is placed in a clean container (4) to collect the filtrate; the filtration device (3) should be sealed to prevent evaporation of the sample moisture.
(S4):將200mL的超純水倒入過濾杯(31)中,調節磁石力攪拌機(32)轉速大於100rpm〔一般不超過200rpm,例如120rpm〕。同時打開閥門通入高純氮氣(5),調節氣體壓力為0.5MPa,進行約0.5-1h的薄膜預壓密。(S4): 200 mL of ultrapure water is poured into the filter cup (31), and the rotation speed of the magnet force mixer (32) is adjusted to be greater than 100 rpm (generally not exceeding 200 rpm, for example, 120 rpm). At the same time, the valve is opened to pass high-purity nitrogen gas (5), the gas pressure is adjusted to 0.5 MPa, and the film pre-compacting is performed for about 0.5-1 h.
(S5):薄膜預壓密結束後,倒出殘餘的超純水。使用量筒準確量取一定體積〔300-500mL〕經0.22或0.45μm的微濾膜過濾後之水樣,倒入過濾杯(31)中,調節磁石攪拌機(32)轉速大於100rpm〔一般為120rpm〕。同時打開閥門通入高純氮氣(5),調節氣體壓力為0.5MPa。(S5): After the film is pre-compacted, the residual ultrapure water is poured out. Use a measuring cylinder to accurately measure a certain volume [300-500 mL] of the water sample filtered through a 0.22 or 0.45 μm microfiltration membrane, pour into a filter cup (31), and adjust the rotation speed of the magnet mixer (32) to be greater than 100 rpm. It is 120 rpm]. At the same time, the valve is opened to pass high-purity nitrogen (5), and the gas pressure is adjusted to 0.5 MPa.
(S6):待收集的濾液體積約為250mL時,停止通入氮氣(5)。準確測定濾液體積,以計算濃縮液的體積,之後可求得對原水的濃縮倍數。將濃縮液倒入乾淨玻璃瓶中進行保存。(S6): When the volume of the filtrate to be collected is about 250 mL, the nitrogen gas (5) is stopped. The volume of the filtrate is accurately determined to calculate the volume of the concentrate, after which the concentration factor for the raw water can be determined. Pour the concentrate into a clean glass bottle for storage.
而上述之S2步驟到S6步驟為樣品的奈米薄膜預處理過程;接續以下步驟:The above steps S2 to S6 are the nano film pretreatment process of the sample; the following steps are carried out:
(S7):對濃縮液進行相關水質分析,包含總氮(採用德國liquidTOC進行總氮〔TN〕檢測)、硝酸鹽氮及亞硝酸氮〔NO3 - 及NO2 - ,採用美國戴安DANEOX離子層析儀測定〕、氨氮〔NH4 + ,採用比色法,並利用UNICO紫外分光光度計,分別在波長420nm及540nm下進行量測〕。(S7): Correlation water quality analysis of concentrated liquids, including total nitrogen (using German liquidTOC for total nitrogen [TN] detection), nitrate nitrogen and nitrite nitrogen [NO 3 - and NO 2 - , using American Dian's DANEOX ion The chromatographic measurement], ammonia nitrogen [NH 4 + , using a colorimetric method, and measurement using a UNICO ultraviolet spectrophotometer at wavelengths of 420 nm and 540 nm, respectively).
(S8):將測定的數值進行差量計算,即DON=TN-NO3 - -NH4 + -NO2 - ,將所得的DON濃度除以濃縮倍數,即可求出為原水的DON濃度。(S8): The measured value is subjected to a difference calculation, that is, DON=TN-NO 3 - -NH 4 + -NO 2 - , and the obtained DON concentration is divided by the concentration factor to obtain the DON concentration of the raw water.
為使本發明之方法更加清晰了解其測定過程與最終計算出的數值差異,特列舉下列諸等實際檢測之實施案例作為說 明:In order to make the method of the present invention more clearly understand the difference between the measurement process and the final calculated value, the following examples of practical detection are specifically mentioned as saying Bright:
以某自來水廠清水經0.22或0.45μm微濾膜過濾後之水樣為待測水樣,使用HCl和NaOH調整pH至中性,並採用不同類型的奈米薄濾膜進行DON測試的預處理;控制水樣的濃縮倍數為6倍左右,對比3種類型奈米薄膜對測定結果的影響。由第四圖所示,在未做奈米過濾預處理條件下,同一個水樣經由差量法直接測定水樣中DON濃度,4次重複量測的濃度變化範圍為-0.1-+0.4mg/L,標準偏差為0.21mg/L,完全無法得到有效的DON濃度數值;經由不同奈米薄膜過濾預處理後,測定DON濃度的標準偏差均顯著減少,有效提高了測定結果的再現性。由於不同類型的奈米薄膜具有不同的有機物截留和吸附性質,這直接影響到DON的回收率高低,導致不同類型奈米薄膜最終測定的DON值略有不同。經計算後,奈米薄膜3的有機物回收率最高,可達到80-95%,因此具有最高的DON測定值。The water sample filtered by a tap water mill with 0.22 or 0.45 μm microfiltration membrane is used as the water sample to be tested. The pH is adjusted to neutral using HCl and NaOH, and different types of nanofiltration membranes are used for DON test. The pretreatment; controlling the concentration of the water sample is about 6 times, comparing the effects of the three types of nanofilm on the measurement results. As shown in the fourth figure, under the condition of no nanofiltration pretreatment, the same water sample directly measures the DON concentration in the water sample by the difference method, and the concentration range of the four repeated measurements is -0.1-+. 0.4mg/L, the standard deviation was 0.21mg/L, and the effective DON concentration value could not be obtained at all. After pretreatment with different nano membranes, the standard deviation of DON concentration was significantly reduced, which effectively improved the measurement results. Reproducibility. Since different types of nano-films have different organic matter retention and adsorption properties, this directly affects the recovery rate of DON, resulting in slightly different DON values for different types of nano-films. After calculation, the nano-film 3 has the highest organic recovery rate of 80-95%, and thus has the highest DON measurement value.
以某自來水廠的清水經0.22或0.45μm的微濾膜過濾後水樣為待測水樣,研究水樣的pH在3-10之間,對本預處理方法DON測定濃度的影響。試驗中採用奈米薄膜3,控制水樣的濃縮倍數為6倍,如第五圖所示,為顯示不同pH條件下,DON的測定結果以及DIN/TDN比值變化, 由圖中可知,pH在6.5-7.5時DON的濃度測定值最高,其原因為:由於不同PH值下,奈米薄膜對水中的無機離子的截留效果不同,最終導致DIN/TDN的比重有所不同;在pH=5.0-9.0時水中DIN/TDN含量維持在60%左右,基本不變;除此之外DIN/TDN都會過高,這將導致對DON測定不準確。After the water of a water plant is filtered through a 0.22 or 0.45 μm microfiltration membrane, the water sample is the water sample to be tested, and the pH of the water sample is between 3 and 10, which affects the concentration of the pretreatment method DON. In the test, the nano film 3 was used, and the concentration ratio of the control water sample was 6 times. As shown in the fifth figure, the DON measurement result and the DIN/TDN ratio change were displayed under different pH conditions. It can be seen from the figure that the concentration of DON is the highest when the pH is between 6.5 and 7.5. The reason is that the interception effect of the inorganic film on the inorganic ions in the water is different due to different pH values, which eventually leads to DIN/TDN. The specific gravity is different; the DIN/TDN content in the water is maintained at about 60% at pH=5.0-9.0, which is basically the same; otherwise, the DIN/TDN will be too high, which will result in the determination of DON. accurate.
以某自來水廠出廠水經0.22或0.45μm的微濾膜過濾後水樣為待測水樣,試驗採用調整不同的過濾水樣體積,保持固定濃縮液體積〔50cm3 〕的方法,控制待測水樣總體積與奈米薄膜過濾膜面積比為2.5、5、7.5、12.5及25cm3 /cm2 ,其對應濃縮倍數分別為2、4、6、10及20,並對濃縮液之DON進行分析比較,結果如第六圖所示,由圖中可以看出,待測水樣總體積與奈米薄膜過濾膜面積比控制在7.5-12.5cm3 /cm2 ,對應之濃縮倍數為6-10倍時,所得濃縮液中DIN比例最低、且DON含量最高。濃縮倍數過高時會導致明顯的濃度極化現象,故欲測定之DON穿透奈米薄膜的量明顯增加,而導致量測DON濃度明顯降低。After the water from a tap water factory is filtered through a 0.22 or 0.45 μm microfiltration membrane, the water sample is the water sample to be tested. The test uses a method of adjusting the volume of the filtered water sample to maintain the volume of the fixed concentrate [50 cm 3 ]. The ratio of the total volume of the water sample to be tested to the area of the nanofiltration membrane is 2.5, 5 , 7.5, 12.5 and 25 cm 3 /cm 2 , and the corresponding concentration multiples are 2 , 4, 6, 10 and 20, and the analysis and comparison of the DON of the concentrated liquid, the results are shown in the sixth figure, as can be seen from the figure, the ratio of the total volume of the water sample to be tested and the area of the nanofiltration membrane is controlled at 7.5-12.5 cm. 3 /cm 2 , corresponding to the concentration factor of 6-10 times, the obtained concentrate has the lowest DIN ratio and the highest DON content. When the concentration factor is too high, it will cause obvious concentration polarization, so the amount of DON penetrating nano film to be measured is obviously increased, and the measured DON concentration is significantly reduced.
對上海市某水淨水廠進行4次不同採樣日期、各處理單元過程中DON濃度的變化進行量測分析,並對直接量測和經由奈米薄膜過濾預處理後再進行測量的DON濃度比較,結果 顯示於第七圖中。從第七圖可以發現,未做奈米薄膜過濾預處理的待測水中DON無法測定,因為高DIN/TDN的干擾,導致TN與DIN分析誤差大,使得在預臭氧與沉澱過程中計算之DON平均濃度成為負值,且量測結果的標準偏差均偏高,無法由該分析方法了解DON在淨水處理過程中的變化。採用奈米薄膜預處理後的分析結果良好,所有測定DON均為正值,且重複試驗標準偏差均在10%以內。經由本方法,可了解DON在飲用水處理過程各單元中的變化情形與溶解性有機碳〔DOC〕的變化趨勢相同,且與各處理單元的特性相符。Measured and analyzed the changes of DON concentration in four different sampling dates and treatment units in a water purification plant in Shanghai, and compared the direct measurement and the DON concentration measured by pretreatment with nano membrane filtration. ,result Shown in the seventh picture. From the seventh figure, it can be found that the DON in the water to be tested which has not been pretreated by nano film filtration cannot be measured. Because of the high DIN/TDN interference, the error of TN and DIN analysis is large, so that DON is calculated in the pre-ozone and precipitation process. The average concentration becomes negative, and the standard deviation of the measurement results are high. It is impossible to understand the change of DON in the process of water purification by the analysis method. The analysis results after pretreatment with nanofilm were good, all DONs were positive, and the standard deviation of repeated tests was within 10%. Through this method, it can be understood that the change of DON in each unit of drinking water treatment process is the same as that of dissolved organic carbon [DOC], and is consistent with the characteristics of each treatment unit.
本發明分析測定方法在進行低濃度DON濃縮的同時,可有效降低濃縮液中無機氮〔含硝酸鹽、亞硝酸鹽和氨氮〕在總氮中的比例,進而大大提高DON測試的精確度和準確度,克服了直接水樣測定法中DON經常出現負值、濃度標準偏差大等難題;本檢測方法測定過程簡單、迅速有效、且具有易於操作、成本低,可廣泛應用地表和地下水監測、大、中和小型淨水廠與污水廠的DON監測和相關研究分析。The analytical method of the invention can effectively reduce the proportion of inorganic nitrogen (containing nitrate, nitrite and ammonia nitrogen) in the total nitrogen while carrying out the concentration of DON at a low concentration, thereby greatly improving the accuracy and accuracy of the DON test. Degree, overcomes the problem that DON often has negative values and large standard deviation of deviation in direct water sample determination method; the detection method is simple, rapid and effective, and has easy operation and low cost, and can be widely used for surface and groundwater monitoring, large DON monitoring and related research and analysis of medium and small water purification plants and sewage treatment plants.
以上所舉者僅係本發明之部份實施例,並非用以限制本發明,致依本發明之創意精神及特徵,稍加變化修飾而成者,亦應包括在本專利範圍之內。The above is only a part of the embodiments of the present invention, and is not intended to limit the present invention. It is intended to be included in the scope of the present invention.
藉由以上所述,該元件之組成與使用實施說明可知,本發明與現有技術相較下,係具有諸多優點,敘述如下:From the above, the composition and use of the components show that the present invention has many advantages compared with the prior art, and is described as follows:
1、本發明測定水中低濃度溶解性有機氮濃度的方 法,可有效降低濃縮後水樣中無機氮〔硝酸鹽氮、亞硝酸鹽氮和氨氮〕濃度占總溶解性有機氮濃度中的比例,同時提高溶解性有機氮的含量,以大大降低利用差量法計算有機氮含量時可能造成的濃度偏差,進而確保量測濃度的精確性和準確性。1. The method for determining the concentration of dissolved organic nitrogen in water at low concentration according to the present invention The method can effectively reduce the concentration of inorganic nitrogen [nitrate nitrogen, nitrite nitrogen and ammonia nitrogen] in the concentration of total dissolved organic nitrogen in the concentrated water sample, and increase the content of dissolved organic nitrogen, so as to greatly reduce the utilization difference. The concentration deviation that may be caused by the calculation of the organic nitrogen content, thereby ensuring the accuracy and accuracy of the measured concentration.
2、本發明與國際上提出的透析膜預處理法相比,預處理時間可至少節省5倍以上,超純水使用量可至少節省720倍以上,且本發明操作簡單、控制方便、經濟可行性極高。2. Compared with the internationally proposed dialysis membrane pretreatment method, the pretreatment time can be saved by at least 5 times, and the ultrapure water usage can be saved by at least 720 times, and the invention has simple operation, convenient control and economic feasibility. Extremely high.
綜上所述,本發明實施例確能達到所預期之使用功效,又其所揭露之具體構造,不僅未曾見諸於同類產品中,亦未曾公開於申請前,誠已完全符合專利法之規定與要求,爰依法提出發明專利之申請,懇請惠予審查,並賜准專利,則實感德便。In summary, the embodiments of the present invention can achieve the expected use efficiency, and the specific structure disclosed therein has not been seen in similar products, nor has it been disclosed before the application, and has completely complied with the provisions of the Patent Law. And the request, the application for the invention of a patent in accordance with the law, please forgive the review, and grant the patent, it is really sensible.
(2)‧‧‧奈米薄膜(2)‧‧‧Nano film
(3)‧‧‧過濾設備(3) ‧‧‧Filter equipment
(31)‧‧‧過濾杯(31)‧‧‧Filter Cup
(32)‧‧‧磁石力攪拌機(32)‧‧‧Magnetic mixer
(4)‧‧‧容器(4) ‧ ‧ container
(5)‧‧‧氮氣(5) ‧‧‧nitrogen
Claims (10)
Priority Applications (1)
Application Number | Priority Date | Filing Date | Title |
---|---|---|---|
TW99126923A TWI409448B (en) | 2010-08-12 | 2010-08-12 | Process for measurements of concentration of dissolved organic nitrogen (don) with low concentration in water |
Applications Claiming Priority (1)
Application Number | Priority Date | Filing Date | Title |
---|---|---|---|
TW99126923A TWI409448B (en) | 2010-08-12 | 2010-08-12 | Process for measurements of concentration of dissolved organic nitrogen (don) with low concentration in water |
Publications (2)
Publication Number | Publication Date |
---|---|
TW201207374A TW201207374A (en) | 2012-02-16 |
TWI409448B true TWI409448B (en) | 2013-09-21 |
Family
ID=46762174
Family Applications (1)
Application Number | Title | Priority Date | Filing Date |
---|---|---|---|
TW99126923A TWI409448B (en) | 2010-08-12 | 2010-08-12 | Process for measurements of concentration of dissolved organic nitrogen (don) with low concentration in water |
Country Status (1)
Country | Link |
---|---|
TW (1) | TWI409448B (en) |
Families Citing this family (3)
Publication number | Priority date | Publication date | Assignee | Title |
---|---|---|---|---|
CN103969411B (en) * | 2014-04-09 | 2015-12-30 | 陈白杨 | A kind of method directly measuring soluble organic nitrogen concentration in water body based on electrodialysis pretreatment mode |
CN104266892A (en) * | 2014-10-29 | 2015-01-07 | 中国烟草总公司湖北省公司 | Method for testing total nitrogen in tobacco by using hydrogen peroxide instead of mercuric oxide |
CN111965374B (en) * | 2020-08-18 | 2023-10-17 | 江西师范大学 | Automatic equipment for detecting content of multistage acidic polysaccharide in water body and application method of automatic equipment |
-
2010
- 2010-08-12 TW TW99126923A patent/TWI409448B/en not_active IP Right Cessation
Non-Patent Citations (2)
Title |
---|
Chakkrid Sattayatewa, "Dissolved organic nitrogen in wastewater effluents:occurrence, fate, and bioavailability", Environmental Engineering in the Graduate College of the Illinois Institute of Technology, 2008 * |
Wontae Lee et al., "Dissolved Organic Nitrogen Measurement Using Dialysis Pretreatment", Environ. Sci. Technol., Vol. 39, pp. 879-884, 2005 * |
Also Published As
Publication number | Publication date |
---|---|
TW201207374A (en) | 2012-02-16 |
Similar Documents
Publication | Publication Date | Title |
---|---|---|
Sim et al. | A review of fouling indices and monitoring techniques for reverse osmosis | |
CN101782566B (en) | Method for measuring concentration of low-concentration soluble organic nitrogen in water | |
Schäfer et al. | Natural organic matter removal by nanofiltration: effects of solution chemistry on retention of low molar mass acids versus bulk organic matter | |
Braeken et al. | Regeneration of brewery waste water using nanofiltration | |
Wei et al. | Size and resin fractionations of dissolved organic matter and trihalomethane precursors from four typical source waters in China | |
Huang et al. | Effects of magnetic ion exchange pretreatment on low pressure membrane filtration of natural surface water | |
Huang et al. | Novel approach for the analysis of bench-scale, low pressure membrane fouling in water treatment | |
Alhadidi et al. | Effect of testing conditions and filtration mechanisms on SDI | |
Myat et al. | Characterisation of organic matter in IX and PACl treated wastewater in relation to the fouling of a hydrophobic polypropylene membrane | |
WO2018133378A1 (en) | Dialysis pretreatment-based method for measuring concentration of dissolved organic nitrogen in municipal sewage | |
AU2015230864A1 (en) | Method for measuring concentration of dissolved organic nitrogen of secondary effluent | |
Fujioka et al. | Integrity of reverse osmosis membrane for removing bacteria: New insight into bacterial passage | |
CN103969411B (en) | A kind of method directly measuring soluble organic nitrogen concentration in water body based on electrodialysis pretreatment mode | |
TWI409448B (en) | Process for measurements of concentration of dissolved organic nitrogen (don) with low concentration in water | |
WO2017202029A1 (en) | Method and system for measuring content of dissolved organic halogens in water | |
Vera et al. | Monitoring UF membrane performance treating surface-groundwater blends: limitations of FEEM-PARAFAC on the assessment of the organic matter role | |
Łaskawiec et al. | The use of membrane techniques in swimming pool water treatment | |
Guo et al. | Ultrafiltration performance of EfOM and NOM under different MWCO membranes: Comparison with fluorescence spectroscopy and gel filtration chromatography | |
Lozier et al. | Microbial removal and integrity monitoring of high-pressure membranes | |
CN104147937B (en) | A kind of detection method of integrity of filtration membranes | |
CN206219360U (en) | A kind of purification detection supply system of drinking water | |
Lipp et al. | Characterization of nanoparticulate fouling and breakthrough during low-pressure membrane filtration | |
Liu et al. | Pressure-driven membrane desalination | |
CN107219102A (en) | A kind of method of soluble organic nitrogen concentration in superposition of electric field pretreatment mode direct measurement water body based on dialysis | |
Voigtländer | Comparison and optimization of removal of natural organic carbon from raw water with ultrafiltration in pilot scale experiments |
Legal Events
Date | Code | Title | Description |
---|---|---|---|
MM4A | Annulment or lapse of patent due to non-payment of fees |