Disclosure of Invention
The invention aims to provide a detection method for degrading impurities in mizoribine bulk drug, which has good impurity separation degree, and can accurately position specific impurities in a chromatogram under a liquid phase condition and identify impurity structures.
In order to achieve the above purpose, the technical scheme adopted by the invention is as follows:
the detection method of the degradation impurity in the mizoribine bulk drug adopts liquid-mass spectrometry to determine the content and the structure of the degradation impurity in the mizoribine bulk drug, and comprises the following steps:
(1) Adopting two-dimensional liquid phase to carry out content measurement, trapping and desalination on degradation impurities in the mizoribine bulk drug;
(2) And (3) adopting mass spectrometry to carry out structural measurement on the degradation impurities trapped in the step (1).
The technical scheme of the invention is further improved as follows: the step (1) specifically comprises the following steps:
s1: preparing a test solution: taking a mobile phase as a solvent, dissolving and fixing the volume of the mizoribine raw material medicine to obtain a sample solution;
s2: preparing a reference substance liquid: diluting the sample solution with mobile phase as diluent to obtain reference solution;
s3: and (3) measuring: precisely measuring the solution of the sample and the solution of the reference substance respectively, injecting into a two-dimensional liquid chromatograph, and recording the chromatograms under the following specific conditions:
1 st -D liquid chromatography conditions: chromatographic column: c18 column (4.6 mm. Times.250 mm,5 μm), mobile phase: phosphoric acid in water-methanol (98:2), flow rate: 0.8ml/min, column temperature: 20 ℃, detection wavelength: 220nm, sample injection amount: 20. Mu.L;
2 nd -D liquid chromatography conditions: chromatographic column: c18 column (2.1 mm. Times.50 mm,2 μm), detection wavelength: 220nm, column temperature: 20 ℃, flow rate: sample injection amount of 0.3 ml/min: 20 μl, mobile phase a:0.1% formic acid, mobile phase B: acetonitrile, gradient elution procedure: 0-3.00 minutes, 95% mobile phase A and 5% mobile phase B; 3.00-8.00 minutes, the mobile phase A gradually changes from 95% to 70%, and the mobile phase B gradually changes from 5% to 30%;8.00-10.00 minutes, 70% mobile phase A and 30% mobile phase B; 10.00-10.10 minutes, the mobile phase A gradually changes from 70% to 95%, and the mobile phase B gradually changes from 30% to 5%; the mobile phase A is 95% and the mobile phase B is 5% in 10.10-12.00 minutes.
The technical scheme of the invention is further improved as follows: the mass spectrum conditions of the step (2) are as follows: ion source: ESI (+/-), interface voltage: 4.5kV, scanning range: primary mass spectrometry: m/z is 200-1500; automatic multistage: m/z is 100-1000; atomizing gas: nitrogen 1.5L/min, dry gas: nitrogen 10L/min, collision gas: argon, desolventizing tube temperature: 200 ℃, heating module temperature: 200 ℃, detector voltage: 1.65kV, calibration method: and (5) automatically tuning the optimized voltage and calibrating the mass number by an external standard method.
The technical scheme of the invention is further improved as follows: the preparation process of the phosphoric acid aqueous solution of the mobile phase is to dissolve 1ml of phosphoric acid in 1500ml of purified water.
The technical scheme of the invention is further improved as follows: said 1 st the-D liquid chromatography column is Venusil MP C18 chromatography column, 2 nd the-D liquid chromatographic column is InertSustatin TM C18 chromatographic column
By adopting the technical scheme, the invention has the following technical effects:
the detection method has good impurity separation degree, is beneficial to protecting a chromatographic column, and can accurately position specific degradation impurities in a chromatogram under the condition of liquid phase of the mizoribine bulk drug and identify an impurity structure.
The invention completes the desalination process of the mobile phase while rapidly analyzing the target peak, and the detection method is simple, convenient and reliable.
The invention is described in 1 st And methanol is added into the mobile phase D, so that the solubility of a sample is ensured under the corresponding liquid chromatography condition, the impurity separation degree is good, the detection result is accurate, and the protection of a chromatographic column is facilitated.
Detailed Description
The invention is further described in detail below with reference to the attached drawings and specific examples:
the detection method of the degradation impurity in the mizoribine bulk drug adopts liquid-mass spectrometry to determine the content and the structure of the degradation impurity in the mizoribine bulk drug, and comprises the following steps:
(1) The method adopts two-dimensional liquid phase to carry out content measurement, trapping and desalination on degradation impurities in the mizoribine bulk drug, and specifically comprises the following steps:
s1: preparing a test solution: taking a mobile phase as a solvent, dissolving and fixing the volume of the mizoribine raw material medicine to obtain a sample solution;
s2: preparing a reference substance liquid: diluting the sample solution with mobile phase as diluent to obtain reference solution;
s3: and (3) measuring: precisely measuring the solution of the sample and the solution of the reference substance respectively, injecting into a two-dimensional liquid chromatograph, and recording the chromatograms under the following specific conditions:
1 st -D liquid chromatography conditions: chromatographic column: venusil MP C18 column (4.6 mm. Times.250 mm,5 μm), mobile phase: phosphoric acid aqueous solution-methanol (98:2) prepared by dissolving 1ml phosphoric acid in 1500ml purified water at a flow rate: 0.8ml/min, column temperature: 20 ℃, detection wavelength: 220nm, sample injection amount: 20. Mu.L;
2 nd -D liquid chromatography conditions: inertSustain TM C18 column (2.1 mm. Times.50 mm,2 μm), detection wavelength: 220nm, column temperature: 20 ℃, flow rate: sample injection amount of 0.3 ml/min: 20 μl, mobile phase a:0.1% formic acid, mobile phase B: acetonitrile, gradient elution procedure: 0-3.00 minutes, 95% mobile phase A and 5% mobile phase B; 3.00-8.00 minutes, the mobile phase A gradually changes from 95% to 70%,mobile phase B was graded from 5% to 30%;8.00-10.00 minutes, 70% mobile phase A and 30% mobile phase B; 10.00-10.10 minutes, the mobile phase A gradually changes from 70% to 95%, and the mobile phase B gradually changes from 30% to 5%; the mobile phase A is 95% and the mobile phase B is 5% in 10.10-12.00 minutes.
(2) And (3) adopting mass spectrum to carry out structural measurement on the degradation impurities trapped in the step (1), wherein the mass spectrum conditions are as follows: ion source: ESI (+/-), interface voltage: 4.5kV, scanning range: primary mass spectrometry: m/z is 200-1500; automatic multistage: m/z is 100-1000; atomizing gas: nitrogen 1.5L/min, dry gas: nitrogen 10L/min, collision gas: argon, desolventizing tube temperature: 200 ℃, heating module temperature: 200 ℃, detector voltage: 1.65kV, calibration method: and (5) automatically tuning the optimized voltage and calibrating the mass number by an external standard method.
Example 1
1. Instrument and reagent
Shimadzu 2D-LC-IT-TOF-MS/MS,
mizoribine (lot numbers 1906002, 20190901),
phosphoric acid (national instruments for medicine, analytical grade), methanol (Fisher, USA), purified water was prepared in the laboratory.
2. Chromatographic conditions
1 st -D liquid chromatography conditions: venusil MP C18 column (4.6 mm. Times.250 mm,5 μm); taking phosphoric acid aqueous solution-methanol (98:2) as a mobile phase, wherein the preparation process of the phosphoric acid aqueous solution comprises the steps of dissolving 1ml of phosphoric acid into 1500ml of purified water; the flow rate is 0.8ml/min, the column temperature is 20 ℃, the detection wavelength is 220nm, and the sample injection amount is as follows: 20. Mu.L.
2 nd -D liquid chromatography conditions: inertSustain TM C18 column (2.1 mm. Times.50 mm,2 μm), detection wavelength: 220nm, column temperature: 20 ℃, flow rate: 0.3ml/min, sample injection amount of 20 μl, mobile phase: 0.1% formic acid-acetonitrile, gradient elution, elution procedure see table below (valve switch flow path into mass spectrum after 1.5 min):
time (min)
|
0
|
3.00
|
8.00
|
10.00
|
10.10
|
12.00
|
Mobile phase a (%)
|
95
|
95
|
70
|
70
|
95
|
95
|
Mobile phase B (%)
|
5
|
5
|
30
|
30
|
5
|
5 |
Mass spectrometry conditions: ion source: ESI (+/-), interface voltage: 4.5kV, scanning range: primary mass spectrometry: m/z is 200-1500; automatic multistage: m/z is 100-1000; atomizing gas: nitrogen 1.5L/min, dry gas: nitrogen 10L/min, collision gas: argon, desolventizing tube temperature: 200 ℃, heating module temperature: 200 ℃, detector voltage: 1.65kV, calibration method: and (5) automatically tuning the optimized voltage and calibrating the mass number by an external standard method.
3. Measurement
Preparation of test solution: taking 50mg of mizoribine bulk drug, precisely weighing, placing into a 25mL volumetric flask, adding a mobile phase (phosphoric acid aqueous solution-methanol (98:2)) to fix the volume and dissolving to prepare a solution containing about 2mg of mizoribine per 1mL, and taking the solution as a sample solution.
Preparation of control solution: precisely measuring 1ml of the sample solution, placing in a 100ml volumetric flask, diluting to scale with mobile phase, and shaking; then, 5ml of the diluted solution was precisely removed and placed in a 50ml volumetric flask to prepare a solution containing about 2. Mu.g per 1ml as a control solution.
And (3) measuring: the samples were measured in an amount of 25. Mu.l, and the samples were injected into a liquid chromatograph to record a chromatogram. The relative main peak retention time of the chromatogram of the sample solution is 0.34, and the impurity peak area is not larger than the main peak area (0.1%) of the control solution, if the impurity peak area is larger than the main peak area of the control solution, the structure identification is needed.
4. Methodological verification
4.1 specificity
Taking the mobile phase as a blank sample, precisely measuring 25 μl, injecting into a liquid chromatograph, and recording the chromatogram. As can be seen from fig. 3, the blank solvent has no interference to the sample measurement.
4.2 limit of detection
The control solution was measured precisely, diluted gradually with mobile phase, measured precisely at 25. Mu.l, injected into a liquid chromatograph, and the sample concentration at 3 XSNR was the detection limit, which was 100ng/ml, as shown in FIG. 4.
5. Analysis of results
5.1 1 st D-LC analysis results
1 st The main purpose of D analysis is to determine the retention time of target impurities to facilitate entrapment into LOOP for two-dimensional desalination analysis. Sample 1 of mizoribine drug substance influence factor test for 0 day and 10 days st D chromatograms are shown in fig. 1 and 2, and can be seen by comparing fig. 1 and 2: the relative retention time of degraded impurities was 0.34 (i.e., the peak with peak time 2.993min in fig. 2).
The test samples of the influence factors of two different existing factories are detected, and the results are shown in the following table:
sample of
|
Specific impurity (%)
|
Other maximum mono-impurity (%)
|
Impurity sum (%)
|
1906002 (0 days)
|
Not detected
|
0.05
|
0.19
|
20190901 (0 days)
|
Not detected
|
0.06
|
0.21
|
1906002 (10 days)
|
0.16
|
0.06
|
0.39
|
20190901 (10 days)
|
0.17
|
0.07
|
0.42 |
Among them, the specific impurity results of 1906002 batches of samples for 10 days are shown in the following table:
sequence number
|
Retention time (min)
|
Area of
|
Area (%)
|
1
|
2.568
|
1244
|
0.03
|
2
|
2.678
|
2149
|
0.05
|
3
|
2.993
|
7526
|
0.16
|
4
|
3.265
|
536
|
0.02
|
5
|
3.372
|
489
|
0.03
|
6
|
3.721
|
2066
|
0.04
|
7
|
3.899
|
3005
|
0.06 |
5.2 2 nd D-LC and Mass Spectrometry analysis results
Degradation impurity (r.t. = 2.993 min) was switched into LOOP 1, pass 2 nd D-LC and mass spectrometry to obtain multi-stage mass spectrograms under positive ion and negative ion modes.
As shown in FIGS. 5-8, the molecular ion peaks are identical in positive and negative ion modes, the cracking rule is identical, and the degradation impurity has a relative molecular weight of 277.0878 and a possible molecular formula C 9 H 15 N 3 O 7 The structural formula is presumed to be as follows:
comparative example 1
Detecting mizoribine in sample (1906002 batches) of mizoribine bulk drug effect factor test for 10 days by adopting a mizoribine detection method reported in literature (journal of Chinese clinical pharmacology, volume 36, 15) for measuring mizoribine concentration in plasma of patients after kidney transplantation by using high performance liquid chromatography-tandem mass spectrometry, 1 st The D chromatogram is shown in FIG. 9, and no impurity peak was detected.
Comparative example 2
1, detecting a sample (1906002 batches) of mizoribine raw material drug influence factor test for 10 days by adopting a detection method of mizoribine related substances recorded in Japanese pharmacopoeia, wherein a mobile phase is a phosphoric acid aqueous solution and a pure phosphoric acid aqueous solution system is adopted as a mobile phase, and 1 st The D chromatogram is shown in FIG. 10, 2.568, 3.279 andthe peak at 3.813 minutes cannot be effectively separated, and a plurality of impurities are possibly combined to form a peak, so that the single impurity amount is high, the upper limit specified in Japanese pharmacopoeia is reached, and the specific impurity results are shown in the following table: