A kind of qualitative and quantitative analysis SF
6the method of gas discharge decomposition product
Technical field
The invention belongs to analysis detection field, be specifically related to a kind of qualitative and quantitative analysis SF
6the method of gas discharge decomposition product.
Background technology
SF
6gas has the good feature of stable chemical nature, safety non-toxic, insulation and arc extinction performance at normal temperatures, makes it be widely deployed in various electrical equipment as a kind of excellent insulating medium.But, along with the electric equipment operation time increases, various insulation fault problem will be there is, be characterized in before not puncturing, shelf depreciation occurring at device interior insulating medium.
SF
6generation, under the effect of shelf depreciation, is decomposed and is produced subfluoride, the O in these sub-stable materials and equipment by gas
2or H
2o reacts further and generates SO
2f
2, SOF
4, SOF
2, SO
2, H
2s, S
2f
10o, HF, CF
4deng product.Wherein, acidic materials SO
2, H
2s, HF have corrosive attack, destroy material and cause dielectric strength to decline, and then cause industrial accident.
For shelf depreciation, traditional detection method is as serious by on-the-spot electromagnetic interference (EMI) in electrical measuring method, ultrasonic method, and impact judges.By analyzing SF
6the kind of electric discharge decomposition product and concentration, can understand device interior state of insulation, for fault diagnosis provides an effective foundation.At present, research both domestic and external shows, the shelf depreciation that different insulative defect causes, SF
6difference is there is in electric discharge decomposition product in kind and generating rate.Therefore, need to find a kind of effective qualitative and quantitative analysis method, identification SF as much as possible
6electric discharge decomposition product, this is to accurately judging that insulation fault is significant.
At present, SF is analyzed
6the common method of gas discharge decomposition product is vapor-phase chromatography, infra-red sepectrometry.But there are the following problems in actual applications for these methods:
1) analysis result is not comprehensive.Utilize vapor-phase chromatography to SF
6the detection of electric discharge decomposition product, the selection of chromatographic column is even more important, and it determines follow-up qualitative and quantitative analysis to effective separation of target compound.In IEC60480-2004 standard, suggestion adopts Porapak Q packed column to be separated, but as can be seen from (Fig. 1) in feature spectrogram, SO
2f
2retention time and SF
6close, both go out peak simultaneously, can not realize being separated completely.Meanwhile, this pillar can not detect other decomposition products as SOF
4and H
2s.And if request for utilization number is 201010134902.6 " a kind of SF analyzed in electrical equipment
6the method of decomposition product " the Gas Pro chromatographic column of recommending in patent, although can SO be realized
2f
2with SF
6be separated, but SOF
2easily SO is changed in pillar
2and cannot detect, and it is to SOF
4detection also helpless.
2) to be accurately qualitatively not easy.Adopt vapor-phase chromatography to analyze unknown sample, qualitative progress needs retention time and known substance calibrating gas to compare, thus determines unknown sample composition.If jointly go out peak without calibrating gas or some composition, then can cause difficulty to accurately qualitative.
3) quantitatively there is interference mutually.Vapor-phase chromatography can only carry out integral and calculating to complete separation peak shape, and each target compound is separated as failed by testing process completely, and peak shape partial stack overlaps even completely then cannot accurate quantitative analysis, there is interference each other.
4) detection limit is not enough.Infra-red sepectrometry detects SF
6electric discharge decomposition product, although without the need to being separated various component, its detection limit is higher than vapor-phase chromatography, cannot detect the product of lower concentration.
Summary of the invention
Having the object of this invention is to provide one can multianalysis and accurate qualitative, quantitative SF
6the combined gas chromatography mass spectrometry analytical approach of gas discharge decomposition product.
To achieve these goals, technical scheme of the present invention is: provide a kind of qualitative and quantitative analysis SF
6the method of gas discharge decomposition product, key step comprises:
1) sample;
2) utilize gas chromatography-GC-MS to carry out qualitative and quantitative analysis to sample, detecting device is quadrupole rod mass detector, and chromatographic column adopts CP-Sil 5 CB capillary column.
Described step 2) the CP-Sil 5 CB chromatographic column selected, its parameter is: length is 60m, and thickness is 8 μm, and internal diameter is 0.32mm.
Preferably, described step 2) analytical approach, analysis condition is as follows, carrier gas: high-purity He gas (99.9999%); Post case temperature: initial 40 DEG C-60 DEG C, final 150
dEG C-220
dEG C; Heating rate: 5
dEG C/min-20 DEG C/min; Post flow: 0.9mL/min-1.2 mL/min; Purge flow rate: 3 mL/min.
Described step 2) analytical approach, analyze SF
6gas discharge decomposition product, its setting parameter is:
1. chromatographiccondition is as follows:
Carrier gas: high-purity He gas (99.9999%),
Sampling volume: 300 μ L,
Post case temperature: 40
dEG C constant temperature keeps 7min, with 15
dEG C/min ramp to 220
dEG C, keep 1min,
Injector temperature: 200
dEG C,
Sample introduction pattern: shunting,
Split ratio: 10,
Flow control mode: constant linear velocity,
Pressure: 56.1kPa,
Post flow: 1.2mL/min,
Purge flow rate: 3.0mL/min;
2. mass spectrophotometry condition is as follows:
Ion source temperature: 200 DEG C,
Interface temperature: 250
dEG C,
The detector switch time: after starting detection, open at time 2.9min, close at time 4.02min, open at time 4.26min, close at time 7min,
Detecting device acquisition mode: SIM,
Gather quasi-molecular ions: comprise the quasi-molecular ions that mass-to-charge ratio is 69,50,119,44,83,102,105,86,67,48,34,33,64,66.
Described step 2) analytical approach, synchronously utilize retention time and with reference to ion identification target compound, setting retention time deviation range ± 0.069min, setting with reference to Ion Phase to strength variance value 80%.
Described step 2) SF
6the analysis of electric discharge analyte detects and comprises: sample introduction; After starting detection, at SF
6before going out peak, analyze CF
4, C
2f
6(time period 2.9min-4.02min); Treat SF
6go out peak complete, reopen detecting device, Analysis for CO
2, SO
2f
2, SOF
4, SOF
2, H
2s, SO
2(time period 4.26min-7min).
Described step 2) analytical approach, the quantitative employing external standard method of target compound, sets up the typical curve that " concentration-peak area " is corresponding, optional zero crossing in typical curve linear fitting procedure, calculate peak area according to integrating meter, then calculate target compound concentration value by typical curve.For the C that cannot purchase calibrating gas
2f
6, adopt relative C
3f
8correction factor is that 1.0 modes carry out sxemiquantitative.
Described step 2) analytical approach, the quantitative test Monitoring lower-cut of various target compound is as follows:
Table 1 SF
6gas discharge decomposition product quantitative analysis method Monitoring lower-cut
Gas composition |
CF
4 |
CO
2 |
SO
2F
2 |
SOF
4 |
SOF
2 |
H
2S
|
Minimum inspection limit (μ l/l) |
0.6 |
0.1 |
0.3 |
0.4 |
0.2 |
0.6 |
Note: adopt semiquantitative C
2f
6, and detect the SO by chromatographic column remaining influence
2, Monitoring lower-cut cannot be determined.
After the present invention adopts technique scheme, mainly contain following advantage:
1) detected components kind is comprehensive.Single injected sampling, adopts SF
6electric discharge decomposition product analytical approach, can detect CF
4, C
2f
6, CO
2, SO
2f
2, SOF
4, SOF
2, SO
2, H
28 kinds of components such as S.Testing process only needs to complete for 20 minutes, increases work efficiency, and reduces energy consumption.Detect material comprehensive, contribute to fully understanding inside electric appliance state of insulation, accurate failure judgement.
2) to unknown sample fast qualitative.Without the need to buying calibrating gas, utilize the known compound in the ion maps that detects and standard spectrum storehouse to compare can to realize qualitative.Utilize retention time and reference ion relative intensity, in accurate judgement sample, whether target compound exists simultaneously.
3) accurate quantitative analysis.For treating quantification compound, selected good object ion peak, after qualitative confirmation, only needs to carry out integration to collected specific objective quasi-molecular ions, can calculate concentration from typical curve.Even if two kinds even retention time of multiple compounds are close, as long as select different object ion peaks to carry out integral and calculating to different compound, accurate quantitative analysis can be carried out equally.Compared with gas chromatography, eliminate overlap of peaks cause cannot be quantitative impact.
Accompanying drawing explanation
Fig. 1 adopts the Porapak Q packed column of advising in existing IEC60480-2004 standard, carries out the typical color spectrogram that gas chromatographic detection obtains.
Fig. 2 is SF of the present invention in embodiment
6the testing result of gas discharge decomposition product analysis method.
Embodiment
Analysis SF described in the present embodiment
6the gas chromatography-mass spectrography method of gas discharge decomposition product, comprises the following steps:
1) sample.In order to avoid sampling container material is to the absorption detecting sample, the sampler bag that Tedlar material can be adopted to make gathers gaseous sample.During collection, the material of connecting line adopts teflon or stainless steel as far as possible.Before each sampling, utilize vacuum pump to be vacuumized by sampler bag, remove other gaseous impuritieies; Connecting line can carry out purging and vacuumizing with tested gas.The present embodiment adopts certain domestic SF
6the sample gas of gas after 96h shelf depreciation, and people is for adding a certain amount of CF
4, H
2s, with the validity of verification method.
2) Shimadzu gas chromatograph-mass spectrometer (GCMS) (GCMS-QP2010) is adopted to detect sample.Select CP-Sil 5 CB chromatographic column, its parameter is: length is 60m, and thickness is 8 μm, and internal diameter is 0.32mm.If chromatogram column length is short, separating effect is bad.Use high-purity He gas as carrier gas; Pin sample introduction, sampling volume is 300 μ L.
1. chromatographic condition is set:
Post case temperature: 40 DEG C of constant temperature keep 7min, with 15
dEG C/min ramp to 220 DEG C, keep 1min; Injector temperature: 200
dEG C; Sample introduction pattern: shunting, split ratio is 10; Flow control mode: constant linear velocity; Pressure: 56.1kPa; Post flow: 1.2mL/min; Purge flow rate: 3.0mL/min.
2. Mass Spectrometry Conditions is set:
Ion source temperature: 200
dEG C; Interface temperature: 250
dEG C; The detector switch time: after starting detection, open at time 2.9min, close at time 4.02min, open at time 4.26min, close at time 7min, select SIM type collection; Gather quasi-molecular ions: comprise the quasi-molecular ions that mass-to-charge ratio (m/z) is 69,50,119,44,83,102,105,86,67,48,34,33,64,66.
3) qualitative analysis, adopts retention time and with reference to ion identification compound, setting retention time band ± 0.069min, setting reference Ion Phase is to strength variance value 80%.
4) quantitative test, to the characteristic ion peak integration of target compound, substitutes into the peak area obtained " concentration-peak area " typical curve that external standard method draws gained, target compound concentration in calculation sample.
Fig. 2 detects SF for utilizing said method
6the result of gas discharge decomposition product.In figure, the corresponding compound of each quasi-molecular ions is followed successively by CF
4, C
2f
6, CO
2, SO
2f
2, SOF
4, SOF
2, H
2s, SO
2, each content is:
Above disclosedly be only preferred embodiment of the present invention, certainly can not limit the interest field of the present invention with this, therefore according to the equivalent variations that the claims in the present invention are done, still belong to the scope that the present invention is contained.