Summary of the invention
Purpose of the present invention exactly in order to address the above problem, provides a kind of back support loss of weight static equilibrium evaluation training method.
In order to achieve the above object, the present invention has adopted following technical scheme: a kind of back support loss of weight static equilibrium evaluation training method, undertaken by a cover back support loss of weight assessment of function training system, the hardware of this back support loss of weight static equilibrium evaluation training system comprises that training bed, control chamber, main control computer, main display and patient use display device; The angle of inclination of training bed can be regulated, the training bed comprises base, bedstead and bed board, bedstead is installed on the base and links to each other with base is rotatable, is provided with left and right sides pedal assembly at the front end of bedstead, is respectively equipped with the plantar pressure sensing device on the pedal assembly of the left and right sides; Be provided with testing evaluation module and functional training module in the main control computer.Training method comprises testing evaluation and functional training two parts, by the patient being carried out the testing evaluation of keeping the standing balance function under back support loss of weight (or the not loss of weight) form, carry out functional training by the loss of weight of the loss of weight/not balance of keeping to a certain degree of standing.
Described back support loss of weight static equilibrium evaluation training method may further comprise the steps:
A, patient tilt to lie on a bed near bed board, and allow its left and right sides foot step down in respectively on the pedal of the left and right sides, and left and right sides pedal is on the same horizontal plane;
B, regulate the angle of inclination of bed according to patient's practical situation automatically by the doctor, realize loss of weight in various degree, angle is more little, and the loss of weight degree is big more.
C, computer enter the assessment of function system, and according to patient's virtual condition, the doctor carries out the short period initial testing and prepares.
D, patient keep two lower limb dynamic balances, make the patient keep initiatively that center of gravity is mobile within the specific limits highly to be consistent the bar diagram of exerting oneself of left and right sides lower limb by visual feedback, suitably regulate posture and dynamics, keep standing balance.
It is the basis of static posture graph parameter assessment that center of gravity projection in patient's static test process changes, and all will use in described below seven CALCULATION OF PARAMETERS.(center of gratitude is COG) before second rumpbone for general gravity center of human body.Above-mentioned center of gravity projection refers to the projection of gravity center of human body on pedal, because that human body is in the standing balance process side force is very little, so the center of gravity projection of human body can be expressed as:
x=(F
A+F
B-F
C-F
D)·L/G
y=(F
B+F
D-F
A-F
C)·M/G
In the formula, L is the central point of four plantar pressure sensing devices arrives sensing device on X-axis a distance, and M is the central point of four plantar pressure sensing devices arrives sensing device on Y-axis a distance, and G is the patient's human weight behind the loss of weight, F
A, F
B, F
C, F
DBe respectively the reading of four plantar pressure sensing devices; Described normal center of gravity projection zero-bit i.e. the central point of four plantar pressure sensing devices.
The frontal plane hunting frequency of assess patient in the back support loss of weight stands still process changes, the frontal plane hunting frequency has reflected within a certain period of time, center of gravity is projected in the change frequency that frontal plane is the positive negative value of horizontal direction, in algorithm, only represent the speed that the center of gravity projected position changes along the x direction of principal axis, this parameter value is big more, and promptly the frequency of Bian Huaing is high more, and reflection patient shake is serious more, center of gravity control is unstable more, as parkinson disease.
The frontal plane hunting frequency is defined as the projection of patient's center of gravity and moves to right side area and move to the number of times of left field and the ratio of time from right side area from left field, that is:
In the formula, N is the number of times that the center of gravity projection is swung from the left field to the right side area and from the right side area to the left field, the testing time of T for selecting.
Average center of gravity had reflected in the motor process that stands still when assessment back support loss of weight stood still, two lower limb balance of exerting oneself, and absolute value is more little, and the expression balanced degree is good more, on the occasion of big more, shows that right lower extremity is than the more load-bearing of left lower extremity; Negative value is big more, and is then opposite.Clinical data shows that average center of gravity projection value is big more, and body symmetry is poor more, as the apoplexy hemiptegic.It is defined as the ratio of patient's center of gravity projection offset distance sum and times of collection, and computational methods are:
In the formula, Gi is the deviant of each center of gravity projection of gathering with respect to the central point of four sensing devices; I=1,2,3,4 ... N (i, N are integer), N is counting of gathering in the testing time of selecting.
The maximum amplitude of oscillation of frontal plane when assessment back support loss of weight stands still, the two lower limb of the reflection patient maximum of degrees of offset firmly on directions X.This difference is relevant with body symmetry, and difference is big more to show that symmetry is poor, a little less than the stability.Relevant disease such as amputee, the hemiptegic.
The maximum amplitude of oscillation of frontal plane is defined as the ultimate range that the projection of patient's center of gravity is offset normal center of gravity projection zero-bit, and its computational methods are:
Smax=Max|Si|
In the formula, Si is the length that each center of gravity projection of gathering is offset normal center of gravity projection zero-bit, i=1,2,3,4 ... N (i, N are integer).
The average amplitude of oscillation of frontal plane when assessment back support loss of weight stands still under the reflection patient loss of weight state, is shaken severe degree and speed on X-direction in the process that stands still, the big more patient's balanced capacity that shows of its value is poor more, relevant disease such as extrapyramidal disease.Its computing formula is:
In the formula, Si is the length that each center of gravity projection of gathering is offset normal center of gravity projection zero-bit, i=1,2,3,4 ... N (i, N are integer), N is counting of gathering in the testing time of selecting.
Center of gravity motion track total length when assessment back support loss of weight stands still, center of gravity motion track total length has reflected under patient's loss of weight state, center of gravity projection motion track total length when standing still, its value is big more to show that center of gravity projection offset distance summation is long more, patient's balanced capacity is poor more.
Center of gravity projection motion track is divided into centre type, front and back type, left right model, multicenter type and diffusion-type, normally is the multicenter type.The computational methods of multicenter type center of gravity projection motion track total length are:
In the formula, x
i, y
iCoordinate for the center of gravity projection deviation post that collects at every turn.I=0,1,2,3,4 ... N (i, N are integer).
The center of gravity motion track gross area in unit interval when assessment back support loss of weight stands still, the center of gravity motion track gross area has reflected the stability and the balanced degree of health control center of gravity in the unit interval.Setting under the same period, the different rehabilitation training phase is being carried out front and back contrast, determining the training effect, numerical value is big more, and reflection stability and balance are poor more, as parkinson disease, and cerebellar ataxia etc.The computing formula of the center of gravity motion track gross area is in unit interval:
In the formula, x
i, y
iBe projected in the coordinate of the deviation post in the pedal scope for each center of gravity of gathering.I=0,1,2,3,4 ... N (i, N are integer); Tc is each time of gathering.
The translational speed of center of gravity projection when the frontal plane side slew rate when assessment back support loss of weight stands still in the unit interval, frontal plane side slew rate are reflected in the maximum amplitude of oscillation on the X-direction, its value is big more, and it is poor more that the expression center of gravity is controlled.Relevant disease such as torsion spasm, chorea etc.Its computing formula is:
In the formula, S '
MaxBe the previous sampled value of the maximum amplitude of oscillation of frontal plane, T is the sampling time.
Assess good center of gravity percentage ratio, good center of gravity percentage ratio is the time that center of gravity drops in normal center of gravity projection zero-bit ± 5% zone to account for the percentage ratio of whole training times.When patient left and right sides bipod is exerted oneself symmetrically and evenly, if its center is normal center of gravity projection zero-bit, represent center of gravity all to drop on right crus of diaphragm to depart from normal center of gravity projection zero-bit+50%, represent center of gravity all to drop on left foot to depart from normal center of gravity projection zero-bit-50%, center of gravity drop in each zone of normal center of gravity projection zero-bit (± 5%, ± 5% to ± 10%, ± 10% to ± 20%, ± 20% to ± 30%) the computing formula of time ratio as follows:
In the formula, n
3For center of gravity from above-mentioned each value of normal center of gravity projection zero-bit (± 5%, ± 5% to ± 10%, ± 10% to ± 20%, ± 20% to ± 30%) number of times of swing in the scope in the zone, N counts for the interior collection of testing time selected;
Back support loss of weight static equilibrium of the present invention evaluation training method is by carrying out the testing evaluation of keeping the standing balance ability under back support loss of weight (or the not loss of weight) form to the patient, obtain data, determine training time and training strength according to assessment result, targetedly the patient is carried out the training of sleep mode standing balance, help accelerating rehabilitation of patients.
The specific embodiment
Referring to Fig. 1, the back support loss of weight static equilibrium evaluation back support loss of weight static equilibrium that training method adopted evaluation training system of the present invention comprises that training bed 1, control chamber 2, main control computer 3, patient use display device 4 and main display 5.Control chamber 2 is installed on the base 11 of training bed, main control computer 3 is arranged on the operating board (operating board not shown come out), the patient is movably arranged on the next door of training bed with display device 4 and links to each other with the main control computer signal of telecommunication, and main display 5 is arranged on the operating board and with the main control computer signal of telecommunication and links to each other.
Training bed 1 comprises base 11, bedstead 12 and bed board 13, and bedstead 12 is installed on the base 11 and turns by rotating shaft 6 with base 11 and links to each other, and is supported on base 11 tops by two push rods 9 (being separately positioned on both sides).Bed board 13 is fixed on the bedstead 12, is connected with left and right pedal assembly 14 at the front end of bedstead 12, is provided with bed board angle adjusting mechanism 15 between bedstead 12 and base 11, and this bed board angle adjusting mechanism comprises a linear stepping motor.On the base 11 of training bed, a plurality of universal adjustment castors 111 are installed, and correspondingly are provided with a plurality of flexible up and down governor motions 112 of castor that are used to regulate.Bed board angular display unit 8 also is installed on bedstead 12, and bed board angular display unit 8 moves the angle that is used to show bedstead 12 and bed board 13 inclinations with bedstead 12.
Left and right pedal assembly 14 symmetrical configuration are identical, respectively comprise pedal support 141, foot front end buffer 142, the heel axis of guide 143 and pedal 144.Pedal support 141 is fixedly linked with bedstead 12; The foot front end buffer 142 and the heel axis of guide 143 are connected between pedal 144 and the pedal support 141.Around pedal 144, be provided with the plantar pressure sensing device, this plantar pressure sensing device is made up of the cantilever beam that posts four foil gauges, the cantilever beam that posts four foil gauges is inserted in the sole stressed member of being made up of pedal, pedal support, foot front end buffer and the heel axis of guide, and links to each other with the sensing device interface circuit signal of telecommunication in the control chamber 2 respectively.Among Fig. 2, F
A, F
B, F
C, F
DRepresent four plantar pressure sensing devices stressed in the installation site respectively, O is the central point of four plantar pressure sensing devices, the center of gravity projection is placed corresponding four plantar pressure sensing devices (normal center of gravity projection zero-bit overlaps with the central point of four plantar pressure sensing devices) then as the center O of sensing device during with normal person's transfixion.L is the central point of four plantar pressure sensing devices arrives sensing device on X-axis a distance, and M is the central point of four plantar pressure sensing devices arrives sensing device on Y-axis a distance.
Control chamber 2 is installed on the base 11 of training bed, is provided with sensing device interface circuit, motor drive module and power module (not shown come out) in the control chamber, and power module is respectively the sensing device interface circuit and motor drive module provides power supply; Motor drive module links to each other with the linear stepping motor signal of telecommunication in the bed board angle adjusting mechanism.
Main control computer 3 is arranged on the operating board (operating board not shown come out), is provided with data acquisition module, testing evaluation module and functional training module in the main control computer.Data acquisition module links to each other with the sensing device interface circuit signal of telecommunication in the control chamber, and the testing evaluation module links to each other with main display 5 signals of telecommunication with the functional training module.
The patient is movably arranged on the next door of training bed 1 with display device 4 and links to each other with main control computer 3 signals of telecommunication, this patient comprises travelling car 41, is installed in support 42 on the travelling car and rack-mount display screen 43 with display device 4, display screen and support flexibly connect, and the height of display screen and angle can be regulated.
Main display 5 is arranged on the operating board and with main control computer 3 signals of telecommunication and links to each other.
Back support loss of weight static equilibrium evaluation training method of the present invention is undertaken by above-mentioned back support loss of weight static equilibrium evaluation training system.Above-mentioned hardware has been realized automatic loss of weight function, gathers data such as related pressure by sensing device group in the hardware and data acquisition unit simultaneously, and main control computer carries out date processing.The concrete steps of back support loss of weight static equilibrium evaluation training method of the present invention are as follows:
A, patient tilt to lie on a bed near bed board, and allow its left and right sides foot step down in respectively on the pedal of the left and right sides, and left and right sides pedal is on the same horizontal plane.
B, regulate the angle of inclination of bed according to patient's practical situation automatically by the doctor, realize loss of weight in various degree, angle is more little, and the loss of weight degree is big more.
C, enter patient information interface shown in Figure 12, newly-built, revise or read patient information, as day of training, patient number, name, sex, year, body weight, and the patient when just being admitted to hospital symptom---left side paralysis is right paralysed information still; Attention: must choose the patient who needs test or training, otherwise modules such as testing evaluation and functional training can't carry out, simultaneously, system will point out and choose the patient who needs test or training.
D, from computer, access testing evaluation-static equilibrium and select the interface, as shown in figure 13.Automatically show selected numbering and name on this interface, practical situation according to the patient is selected the required testing time by the doctor, the testing time branch stands still 30 seconds, 60 seconds that stood still, free cycle, input loss of weight angle in the yellow negative edit box of loss of weight angle cuts a part of weight of human body automatically to help the patient again.Click " entering test " button then and enter the static equilibrium test interface, as shown in figure 14.
E, as need first patient is tested preparation, click " test is prepared " button, the doctor allows the patient carry out the initial testing of 10 times in second and prepares, and begins to carry out official testing again after allowing the patient of first test adapt to this test.
The patient is according to pressure bar diagram in the left and right sides on patient's display screen, and the two lower limb of active adjustment firmly make left and right sides lower limb pressure bar diagram highly consistent.
F, also can directly click " test beginning ", the left-half of this static equilibrium test interface is provided with the bar diagram of real-time demonstration left and right sides foot to the pressure size of left and right sides pedal, shows selected testing time and loss of weight angle automatically in the selected parameter in the upper right corner; The right half part middle part is every static posture graph parameter viewing area; The lower right corner is provided with " test is prepared ", " test beginning ", " test stops ", " test result ", " typing report " and " returning " six buttons and testing time slider bar, and slider bar shows the testing time progress in real time.
Figure 14 is to be that 60 seconds, loss of weight angle are the static equilibrium test interfaces of 45 degree the testing time.The corresponding 90 kilograms of pressure of the total height of each bar diagram, and calibrate with per 1 kilogram, accurately show the pressure size of left and right sides foot to pedal, by real-time bar diagram, the patient can be clear that the firmly situation of oneself two foot, the doctor can be well understood to the patient body equilibrium condition, being that the patient selects the suitable training project easily.
G, " test stops " functional keys when the patient runs into emergency case, are clicked this button, and test stops, and stopping to test the back data can preserve and show.
H, after arriving selected test period, click " test result " button again, show the test and the assessment result of the parameters of static posture figure on the static test interface automatically.Described parameter comprises frontal plane hunting frequency, average center of gravity, the maximum amplitude of oscillation of frontal plane, the average amplitude of oscillation of frontal plane, center of gravity motion track total length, the center of gravity motion track gross area and frontal plane side slew rate.
I, click " typing report " button, storage patient information and test result data; At this moment,, can see the report of generation, can also report with the word format print if enter reporting modules.
J, click the Back button, leave the static test interface, after obtaining every data of testing evaluation, can power cut-off, also can directly enter functional training.
The method of k, functional training is at first to enter functional training-normal stand balance and select the interface, as shown in figure 15.According to above-mentioned testing evaluation result, can select sleep mode, and input loss of weight angle, click " entering training " button, can enter the normal stand balance---sleep mode training interface, as shown in figure 16.
L, click " training beginning " button are trained, training interface left-half is provided with the picture of real-time demonstration barycenter trajectory, the picture center is normal center of gravity projection zero-bit, represent center of gravity all to drop on right crus of diaphragm to depart from normal center of gravity projection zero-bit+50%, represent center of gravity all to drop on left foot to depart from normal center of gravity projection zero-bit-50%, with the straight line of red, green, blue, Huang Si kind color define respectively depart from normal center of gravity projection zero-bit ± 5%, ± 10%, ± 20% and ± 30% zone.The upper right corner at interface shows the loss of weight angle automatically, and carries out the training time and select, and the training time is divided into half a minute, 1 minute and free time; The right half part middle part is controlled at time ratio and good center of gravity percentage ratio parameter display area in the central area for center of gravity; The lower right corner is provided with " training beginning ", " stopping training ", " checking the result ", " typing report ", " returning " five buttons and training progress slider bar, and training progress slider bar shows the training time progress in real time.
Figure 16 is that the loss of weight angle is the sleep mode training interface that 60 degree, training time is set at 1 minute.Require patient's active adjustment left and right sides lower limb dynamic balance, its center of gravity projected footprint is moved along the picture center, can be clear that the situation of the normal center of gravity projection of the deviation of gravity center zero-bit of oneself by this interface patient.
After m, training finish, click " checking the result " button, show training result on the interface, this result comprise center of gravity drop on from normal center of gravity projection zero-bit ± 5%, ± 5% to ± 10%, ± time ratio in 10% to ± 20% and ± 20% to ± 30% zone and good center of gravity percentage ratio.Good center of gravity percentage ratio is the time that center of gravity drops in normal center of gravity projection zero-bit ± 5% zone to account for the percentage ratio of whole training times, makes things convenient for the doctor not only accurately but also observation all sidedly.
Center of gravity drops on from the value of normal center of gravity projection zero-bit ± 5% big more, shows that patient's states is good more, and ± 10% takes second place, ± 20% more take second place, ± take second place for 30% time.For example be controlled at after twice rehabilitation training of certain person patient ± 5% all be 20% with interior time ratio, and center of gravity is controlled at ± 10% increase with interior percentage of time, then can prove the effectiveness of rehabilitation training equally.
N, click " typing report " button are preserved patient's training result;
O, click the Back button, leave the normal stand balance---sleep mode training interface, finish training or proceed the next round functional training.
The static equilibrium assessment of function training system software realization flow of main control computer is as follows:
At first read patient information: patient number, name, thigh length, shank length etc.; Evaluation time and loss of weight angle are set; Carry out hardware testing (test sensing unit group whether operate as normal, then log off) again as cisco unity malfunction; By lathe bed inclination modular program flow chart, test beginning modular program flow chart, test preparation module program flow diagram, test stopping modular program flow diagram, test result module program flow diagram, typing reporting modules program flow diagram, training beginning modular program flow chart, stop the training module program flow diagram, check that the object module program flow diagram carries out test in proper order.
Lathe bed inclination modular program flow chart as shown in Figure 3.Native system adopts a motion control device, the motor that the corresponding control of wherein (Z axle) of control device lathe bed tilts, and the lifting of lathe bed is that the rotating by motor realizes moving upward and moving downward.Click " entering test " or " entering training " button, just enter lathe bed inclination program circuit, open motion control device, motor movement speed is set.The doctor is according to patient's practical situation input loss of weight angle.Judge whether current lathe bed angle is consistent with the input angle: if the loss of weight angle of doctor's input is greater than current lathe bed angle, calculate the angle that lathe bed need rise, is this angular transition corresponding umber of pulse, control device sends pulse, motor is just changeing the angle of certain umber of pulse correspondence, lathe bed rising certain angle; If equal current lathe bed angle, motor does not change, and lathe bed is motionless; Less than current lathe bed angle, calculate the angle that lathe bed need descend, control device sends pulse, and the motor counter-rotating moves downward the angle of certain umber of pulse correspondence, lathe bed decline certain angle.Check the state of motion control device axis, if busy, the expression motor also is not rotated in place, and continues to check; If not busy, illustrate that motor has rotated to finish, stop the motion of motor, the closing movement control device.
Test beginning modular program flowchart text:
Test beginning module major function is data acquisition and data storage, it designs based on data acquisition unit, and harvester provides built-in function, according to built-in function exploitation harvester class libraries, the everything of management harvester, test beginning modular program flow chart as shown in Figure 4.
Start training evaluation process software, at first carry out the testing evaluation data acquisition and be saved in the file module flow process, click " test beginning " button, activate " test stops " and the Back button, locking " test is prepared ", " test beginning ", " test result " and " typing report " button; The data sampling cycle is set; Obtain the current time, and generate data file (not repeating) to guarantee filename with the current time name; Judge whether patient's file directory exists, then do not create patient's file directory if do not exist, there is the patient's file directory that just directly chdirs, preserve data file to guarantee to forward under patient's file directory, generate the file of expansion .dat by name, form the preservation of patient data files complete name concatenating file and be masked as very, be masked as true respresentation and can preserve data file, be masked as false expression and cannot preserve data file (as " test is prepared ").The doctor sets the testing evaluation time.According to the sampling total data, the application relief area judges whether relief area applies for success, if unsuccessful just return, and prompt facility has problem, if success, setting actual samples data counts variable, and carry out initialization of variable; Transmit patient data files name and file and preserve sign to the harvester class libraries, open data acquisition unit, the voltage sample scope is set, dispose 4 sampling channels, the corresponding left and right plantar pressure sensing device group of difference, sampled data is set preserves the address, open patient data .dat file with WriteMode; The initialization scroll bar; Record static test time started point is set sampling clock; Read all sampling channel data, the discrete sampling channel data: preceding 0,1 passage is left plantar pressure sensing device data, and back 2,3 passages are right plantar pressure sensing device data; The actual samples that adds up in real time data number; To on screen, show in real time with left and right two bar diagrams respectively after the left and right plantar pressure sensing device group data computation, the corresponding 90 kilograms of pressure of the total height of each bar diagram, and calibrate with per 1 kilogram, with the pressure size of left and right sides foot on the time point that shows each collection, import computer by the pressure sensor device that is contained on the pedal to pedal; Write down current point in time once more; Judge whether the testing evaluation time arrive, if the testing time does not arrive, the real-time update scroll bar continues to read the sampling channel data; If the testing time arrives, close sampling clock.Transmit real data total amount and buffer pointer, calculate the testing time, close data acquisition unit; The data of gathering are write patient data files, file is kept on the hard disk close file; The buffer release district, the number variable zero clearing of actual samples data, original state is got back in record static test time started point variable zero clearing.
The explanation of test preparation module program flow diagram:
Test preparation module program flow diagram as shown in Figure 5.Click " test is prepared " button, carry out module flow process shown in Figure 4, lock other buttons, the data sampling cycle is set; Put the file preservation and be masked as vacation, promptly cannot preserve data file; Transmit file and preserve sign to the harvester class libraries, open data acquisition unit, the voltage sample scope is set, configuration sampling channel number: 4 is left and right plantar pressure sensing device group, sampled data is set preserves the address; The initialization scroll bar; Record static test test current point in time is also set sampling clock; Regularly read 4 sampling channel data; Every 50ms shows with left and right two bar diagrams respectively on screen in real time with left and right plantar pressure sensing device data.Write down current point in time once more.The time that test is prepared fixedly was made as 10 seconds, whether judged the testing time by 10 seconds, if less than 10 seconds, the real-time update scroll bar continues to read the sampling channel data; If by 10 seconds, close sampling clock, close data acquisition unit, activate " test is prepared ", " test beginning " and the Back button, locking " test stops ", " test result ", " typing report " button.
The explanation of test stopping modular program flow diagram:
In test process,, then click " test stops " button and finish test this time if the patient does not feel like oneself or other reason need stop test.Test stopping modular program flow diagram as shown in Figure 6.At this moment, program will be carried out module flow process shown in Figure 6, activate " test result ", " test is prepared ", " test beginning " and the Back button, locking " test stops " and " typing report " button.Close sampling clock.Transmit real data total amount and buffer pointer, calculate the testing time, close data acquisition unit; The data of gathering are write patient data files, file is kept on the hard disk close file.The buffer release district, the number variable zero clearing of actual samples data, original state is got back in record static test time started point variable zero clearing.
The explanation of test result module program flow diagram:
The test result module program flow diagram as shown in Figure 7.Click " test result " button, carry out module flow process shown in Figure 7, locking " test stops " and " test result " button activates " test beginning ", " typing report ", " returning " and " test is prepared " button.Judge whether the patient data files name exists, do not return if just do not exist, show and to provide test result, if exist, just open this document with read-only mode, obtain file size, the calculating data number, setting up one is that the relief area of length is used for record data with the file data number, setting up one is that the relief area of length is used to write down position of centre of gravity with 1/4th of file data number, use endless form that the data of file are write the file data relief area, read four channel datas according to preserving sequential loop, the data substitution centroid algorithm of four passages is calculated position of centre of gravity, and center-of-gravity value is put into the center of gravity data buffer zone, up to running through file, close file.Call the average center of gravity in the date processing dynamic link library then, the maximum amplitude of oscillation of frontal plane, the average amplitude of oscillation of frontal plane and barycenter trajectory total length, calculate four static posture graph parameters according to center of gravity data buffer zone center-of-gravity value batch total, call frontal plane side slew rate in the unit interval in the date processing dynamic link library again, the computational algorithm of the interior center of gravity motion track gross area of unit interval and frontal plane hunting frequency during motion, calculate movement time in conjunction with sampling rate, calculate three static posture graph parameters according to center of gravity data buffer zone center-of-gravity value batch total, rule of thumb the data tabular value provides the assessment result of above-mentioned seven parameters.Show the result of calculation and the assessment result of seven static posture graph parameters, discharge all relief areas.
The explanation of typing reporting modules program flow diagram:
Typing reporting modules program flow diagram as shown in Figure 8.Click the typing report button, carry out module flow process shown in Figure 8, locking " test stops ", " test result ", " typing report " button activate other buttons.Judge whether the report file name exists, if do not exist, just regenerate the report file name and preserve the report file name, add patient name and number the report content starting position, report content comprises seven static posture graph parameters: the center of gravity motion track gross area and frontal plane hunting frequency and assessment result in the unit interval when average center of gravity, the maximum amplitude of oscillation of frontal plane, the average amplitude of oscillation of frontal plane, center of gravity motion track total length, frontal plane side slew rate, motion.Judge again whether report file exists,, just create report file, write patient name and numbering and the above-mentioned content for preparing a report if do not exist; If exist, open file with a WriteMode, file pointer moves on to the end, adds the report content of above-mentioned preparation, closes report file then; Typing newspaper module finishes to return.
Said procedure realized with patient same day repeatedly assessment result be kept at function in the identical file, this document can use various software such as Word to check.
Training beginning modular program flowchart text:
Training beginning modular program flow chart as shown in Figure 9.Click " training beginning " button, carry out training start program flow process, this flow process and above-mentioned test start program flow process are basic identical, and difference is date processing and display part.On screen, refresh demonstration according to the left and right lower limb pressure difference value of left and right plantar pressure sensing device group data computation and every 50ms in position of centre of gravity trajectory diagram mode, use the bundle of lines screen of different colours to be divided into ± 5%, ± 10%, ± 20% and ± 30% several intervals, the pressure difference value zero-bit is in the center Screen position, pressure difference value is for negative, promptly show in the zero line leftward position, the expression left foot is big with the force rate right crus of diaphragm, and vice versa; Respectively the real-time statistics pressure difference value drop on ± 5% scope in, ± 5% to ± 10% zone in, ± 10% to ± 20% zone in, ± 20% to ± 30% zone in the quantity and the good center of gravity percentage ratio of point; Wherein good center of gravity percentage ratio is meant the ratio of left and right lower limb pressure difference value in central area ± 5%.
Stop the explanation of training module program flow diagram:
Stop the training module program flow diagram as shown in figure 10.Stop the training module program circuit and above-mentioned test stopping modular program circuit is basic identical.In training process, if the patient feels under the weather or do not think to carry out training this time again and just can click " stopping training " button result and this time train.Click " stopping training " button, activate " checking the result ", " training beginning " and the Back button, locking " stops training " and " typing report " button.
Check the explanation of object module program flow diagram:
Check the object module program flow diagram as shown in figure 11.Click " checking the result " button, locking " stops training " and " checking the result " button, activate " typing report ", " training beginning " and the Back button, calculate left and right lower limb pressure difference value drop in central area ± 5%, drop on central area ± 5% in ± 10%, drop on central area ± 10% in ± 20%, drop on central area ± 20% to ± 30% interior point; Calculate left and right lower limb pressure difference value central area ± 5%, ± 5% to ± 10%, ± 10% to ± 20%, ± percentage ratio in 20% to ± 30%; Show each percent value and good center of gravity control percentage ratio; Show each percent value.
Parameters related in above-mentioned testing evaluation and the functional training is offered relief area according to the testing time that sets automatically by computer, certain hour refreshes once at interval, the cushion space maximum can be deposited the data of certain acquisition time, when test stops, data are kept on the hard disk with file format automatically, so that other modules are calculated and are called, utilize the formula that is placed in advance in the program again, draw various static posture graph parameters by the computer background operation.Formula in the program is all realized by Dynamic link library program, is convenient to upgrade from now on and the content expansion.